Operating temperature     : 0

50

, < 85%RH (without condensation) 

Storage temperature       : 

10

60

, < 80%RH (without condensation) 

Auto power off            : Approx. 10 minutes after power on 

Withstanding voltage            : AC 2000V/1 minute between CT part and grip 

Consumption current            : 14mA (approx. 48h for continuous use) 

Power supply             : AAA alkali battery LR03

×

Dimension/Weight         : 71(W)

×

315(H)

×

37(D)mm, approx. 750g 

Accessories               : Battery LR-03 (installed into the instrument)/3pcs. 

                           Carrying Case/1pce., Instruction Manual/1pce. 

                           Test Lead for voltage input(3 phase)/1set   

 

 

 

Accuracy (23

℃±

5

, less than 85%RH) 

 

Leakage current (Io), Line current (I), Resistive leakage current (Ior), Voltage (V)   

 

Range 

Resolution 

Display 

Accuracy 

I, 

Io 

10mA 

0.001mA 

0~9.999mA 

±

1.0%rdg

±

10dgt 

100mA 

0.01mA 

10.00~99.99mA 

1000mA 

0.1mA 

100.0~999.9mA 

10A 

0.001A 

1.000~9.999A 

Ior 

10mA 

0.001mA 

0~9.999mA 

±

1.5%rdg

±

20dgt 

100mA 

0.01mA 

10.00~99.99mA 

±

1.5%rdg

±

15dgt 

1000mA 

0.1mA 

100.0~999.9mA 

10A 

0.001A 

1.000~9.999A 

500V 

0.1V 

0~499.9V 

±

1.0%rdg

±

10dgt 

 

 

 

The conductor must be located at the center of CT. 

 

 

AC current detection : Dual integration mode 

 

 

The accuracy of Ior is based on the phase angle with 0

°

between the voltage 

and current at the standard voltage of AC100V in case of single phase. In case 

of 3P/3W, it is based on the phase angle with 300

°

between  the  voltage and 

current at 

wiring. 

 

 

For Ior measurement, select 50 or 60Hz by frequency select switch

.

 

 

 

Insulation resistance values are calculated by the formula of voltage and 

current (Ior) at each wiring method. 

 

 

Measuring ranges for Io & Ior are auto but it can be changed to manual, when 

displayed value cannot be read by change of ranges due to big fluctuation of the 

measuring current.    As to the operation, refer to “Change of Current Range” of 

MEASUREMENT. 

 

Содержание MCL-800IR

Страница 1: ...Before you use your new instrument read this instruction manual completely and familiarize yourself thoroughly with all functions and keep this instruction manual carefully to take out whenever you ne...

Страница 2: ...al damage can be supposed caused from mis operations A E A WARNING 2BPOSSIBLE ELECTRICAL SHOCK This instrument is for the use of low voltage circuit Do not make measurements of power lines carrying mo...

Страница 3: ...min resolution 0 001mA Ior 0 10A 4 range auto min resolution 0 01mA V 0 500V 1 range min resolution 0 1V M Calculated by voltage and current 0 10M min resolution 0 001M Input frequency range 50Hz 60Hz...

Страница 4: ...0A 0 001A 1 000 9 999A Ior 10mA 0 001mA 0 9 999mA 1 5 rdg 20dgt 100mA 0 01mA 10 00 99 99mA 1 5 rdg 15dgt 1000mA 0 1mA 100 0 999 9mA 10A 0 001A 1 000 9 999A V 500V 0 1V 0 499 9V 1 0 rdg 10dgt The condu...

Страница 5: ...current and clamp method Open Close Lever CT will open by pushing this lever to inside Power Switch POWER By pressing this switch the power becomes on and pressing once again power off Voltage Displa...

Страница 6: ...correctly By pressing once the circuit setting is 1 and twice is and three times is Y Frequency Switch Select 50Hz or 60Hz according to the circuit to be measured Data Hold D HOLD By pressing one tim...

Страница 7: ...ument B sign will appear on the display when batteries are exhausted and get less than operation voltage Replace to new batteries immediately Do not use the batteries mixed new one and once used and o...

Страница 8: ...ot use the instrument leaving the battery cover off If excessive current is applied to the CT the instrument will be heated and damaged Do not apply more than 50Arms to the CT part Do not apply more t...

Страница 9: ...case of 3 phase 3 Voltage Measurement 500V Range 1 Press POWER switch once All display lightens and the range becomes Io current measurement 2 Press Voltage display switch once and the display changes...

Страница 10: ...the instrument is looking for phase angle of current and voltage Take care for wiring sufficiently Also if frequency setting is not correct the phase angle is changed and the correct measurement canno...

Страница 11: ...switch once Check if the input voltage is correct Press current range switch and select Ior circuit mode to be measured Current range switch will be changed as under by pressing one time 5 To measure...

Страница 12: ...b In case of Single Phase Three Wires Measuring Mode 1 Ior c In case of Three Phase Three Wires Measuring Mode Ior White Red White Red LOAD LOAD Red Red White White Black Black LOAD...

Страница 13: ...large current fluctuation to be measured however the measurement cannot be done In this case fix the current range by manual mode and make measurement Operation of Manual Range 1 Press Power Switch on...

Страница 14: ...ng your name address phone number and defect point WARRANTY This instrument is sent out from our factory after the sufficient internal inspections but if you find any defect due to the fault in our wo...

Отзывы: