SPECIFICATIONS                                                     

 

 
 

1.

 

U

CURRENT DETECTION ZCT 

 

Inside diameter           : 

Φ

80mm 

Method                    : Split core type ZCT 

Withstanding voltage      : AC 2000V/1 minute between CT core and grip 

 

 

 

2.

 

U

MEASURING PART

U

   

 

Measuring function                : Leakage current (Io), Line current (I),   

Resistive leakage current (Ior), Voltage (V), 

Insulation Resistance (M

Ω

){calculated value}   

Measuring method                  : Clamp CT (in case of Ior, based on voltage standard) 

Measuring range          :    Io, I: 0~10A 4 range auto, min. resolution 0.001mA 

                            Ior: 0~10A 4 range auto, min. resolution 0.01mA 

                            V: 0~500V 1 range, min. resolution 0.1V 

                            M

Ω

: Calculated by voltage and current 

                                 0~10M

Ω

, min. resolution 0.001M

Ω

   

Input frequency range        : 50Hz/60Hz (manually by switch) 

AC current detection              : Dual integration mode 

AD conversion             : Successive approximation method 

Display change            : By tact switch 

Display                   : LCD, max.9999 reading with annunciator  

Data hold indication       : “DH” mark on LCD 

Sampling rate             : 2 times/sec. 

Overrange indication      : “OL” mark on LCD 

Low battery indication          : "B" mark on LCD 

Circuit voltage            : less than 600V (isolated wire) 

Auto power off             :Approx. 10 minutes after power on 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Содержание MCL-800IR

Страница 1: ...Before you use your new instrument read this instruction manual completely and familiarize yourself thoroughly with all functions and keep this instruction manual carefully to take out whenever you ne...

Страница 2: ...al damage can be supposed caused from mis operations A E A WARNING 2BPOSSIBLE ELECTRICAL SHOCK This instrument is for the use of low voltage circuit Do not make measurements of power lines carrying mo...

Страница 3: ...min resolution 0 001mA Ior 0 10A 4 range auto min resolution 0 01mA V 0 500V 1 range min resolution 0 1V M Calculated by voltage and current 0 10M min resolution 0 001M Input frequency range 50Hz 60Hz...

Страница 4: ...0A 0 001A 1 000 9 999A Ior 10mA 0 001mA 0 9 999mA 1 5 rdg 20dgt 100mA 0 01mA 10 00 99 99mA 1 5 rdg 15dgt 1000mA 0 1mA 100 0 999 9mA 10A 0 001A 1 000 9 999A V 500V 0 1V 0 499 9V 1 0 rdg 10dgt The condu...

Страница 5: ...current and clamp method Open Close Lever CT will open by pushing this lever to inside Power Switch POWER By pressing this switch the power becomes on and pressing once again power off Voltage Displa...

Страница 6: ...correctly By pressing once the circuit setting is 1 and twice is and three times is Y Frequency Switch Select 50Hz or 60Hz according to the circuit to be measured Data Hold D HOLD By pressing one tim...

Страница 7: ...ument B sign will appear on the display when batteries are exhausted and get less than operation voltage Replace to new batteries immediately Do not use the batteries mixed new one and once used and o...

Страница 8: ...ot use the instrument leaving the battery cover off If excessive current is applied to the CT the instrument will be heated and damaged Do not apply more than 50Arms to the CT part Do not apply more t...

Страница 9: ...case of 3 phase 3 Voltage Measurement 500V Range 1 Press POWER switch once All display lightens and the range becomes Io current measurement 2 Press Voltage display switch once and the display changes...

Страница 10: ...the instrument is looking for phase angle of current and voltage Take care for wiring sufficiently Also if frequency setting is not correct the phase angle is changed and the correct measurement canno...

Страница 11: ...switch once Check if the input voltage is correct Press current range switch and select Ior circuit mode to be measured Current range switch will be changed as under by pressing one time 5 To measure...

Страница 12: ...b In case of Single Phase Three Wires Measuring Mode 1 Ior c In case of Three Phase Three Wires Measuring Mode Ior White Red White Red LOAD LOAD Red Red White White Black Black LOAD...

Страница 13: ...large current fluctuation to be measured however the measurement cannot be done In this case fix the current range by manual mode and make measurement Operation of Manual Range 1 Press Power Switch on...

Страница 14: ...ng your name address phone number and defect point WARRANTY This instrument is sent out from our factory after the sufficient internal inspections but if you find any defect due to the fault in our wo...

Отзывы: