mtts Beluga-V1 Скачать руководство пользователя страница 10

V-0-03-EN

Beluga User manual 

18

Alarm system

Alarm

Cause

Effect

HIGH PRESSURE

Measured pressure higher than the 

PIP setpoint by 4 cmH2O

Device adjusts pressure by itself

Pressure >14 cmH2O for longer 

than 5 seconds

No effect

HIGH FiO2

Measured FiO2 higher than the 

setpoint by 6 %FiO2

Device controls oxygen valve to regulate FiO2

PRESSURE SENSOR FAILURE

Pressure sensor disconnected or 

faulty

No effect

OXYGEN SENSOR FAILURE

Oxygen sensor disconnected or 

faulty

Oxygen valve closed

LOW BATTERY

Battery remaining capacity is low

No effect

List of alarms:

All the alarms listed above are conisdered MEDIUM PRIORITY (ref. IEC 60601-1-8).

In order to verify the operation of auditory and visual alarms, perform the following 

procedure:

1. On the unlocked screen, set PIP to 20 cmH2O

2. Occlude the mixed gas outlet for 10 seconds.

3. Check pressure audible alarm sounds

Alarm system test

General

Cleaning

Comply with hospital, local and national guidelines for product cleaning fre-

quencies.

Ensure oxygen supply is turned off and disconnected from Beluga Ventialator 

before performing cleaning procedures.

Cleaning shall be performed at ambient condictions.

Before cleaning, remove and discard all used disposable products using rec-

ommended method of disposal.

Dust all surfaces with clean damp soft cloth.

Clean all plastic surfaces with mild detergent solution (maximum 2% in wa-

ter).

Dry all the surfaces after cleaning with a clean soft cloth or paper towel.

Ensure that no part of Beluga Ventilator is immersed in any cleaning liquid or 

cleaning solution.

Do not use abrasive cleaning solutions.

Ensure all Beluga Ventilator parts and accessories are checked before re-

turning the device to service.

19

The reusable resuscitation circuit includes a number of parts that must be com-

pletely disassembled and sterilized before use. These parts include:

 Component

  Recommended sterialization method

  Silicone tube

  Autoclaving

  PEEP valve

  Low-temperature plasma sterilization OR

  soak disinfection with 1:500 chlorine-containing disinfectant

  Mask

  Autoclaving

Содержание Beluga-V1

Страница 1: ...Beluga User manual User Manual Effective date March 2020 Issue Number 00 Version 03 EN Issue Date March 2020 www mtts asia com Beluga ventilator Model Beluga V1...

Страница 2: ...r manual Contents Company Information Introduction Device Description Warnings Setup Operation Alarm System Cleaning Maintenance Speci cations Explanation of Symbols Warranty Policy 04 05 06 09 10 12...

Страница 3: ...GY TRANSFER AND SERVICES Co LTD House No 26 Alley 41 An Duong Vuong Street Tay Ho District Hanoi City Vietnam Tel 84 24 3766 6521 Fax 84 24 3718 8050 Email assistance mtts asia com www mtts asia com P...

Страница 4: ...g battery charging conditions 5 Pressure Measurement Displays Measured Pressure 6 FiO2 Measurement Displays Measured FiO2 7 Setting Button Touch this button to enter Setting screen 8 Lock Button Touch...

Страница 5: ...s or nitrous oxide Do not modify this equipment without authorization of the manufacturer Using the device outside of the speci ed ambient temperature range 19 C to 37 C or humidity range 30 RH to 90...

Страница 6: ...device and then directly into the power outlet in the wall B Press the Power Button to turn the device ON Only use T piece circuit and power supply unit provided by MTTS The gas sources must provide p...

Страница 7: ...s B FiO2 Select FiO2 Setpoint circle then touch Up Down button to adjust FiO2 A PIP Select PIP setpoint circle then touch and hold Up Down button to adjust PIP The red horizontal level will move up do...

Страница 8: ...s some features of the device to be con gured To enter Setting screen touch Setting button on the unlocked screen To return the unlocked screen touch the Return button on the Setting screen Resuscitat...

Страница 9: ...ation Oxygen sensor must be calibrated at least every 3 months It is recommended to calibrate the sensor before each use Operation Turning off the device 17 The device can be turned off by Power butto...

Страница 10: ...al and national guidelines for product cleaning fre quencies Ensure oxygen supply is turned off and disconnected from Beluga Ventialator before performing cleaning procedures Cleaning shall be perform...

Страница 11: ...f parts that must be replaced or serviced during the lifetime of the device These parts include Component Approx Duration of Use Maintenance Information Air lter 6 months Clean or replace Oxygen senso...

Страница 12: ...er characteristics Off the shelf external power supply Battery 60W 100 240VAC 47 63Hz International medical safety approvals ANSI AAMI EN 60601 1 UL TUV Class I construction is standard ground require...

Страница 13: ...cturer Date of manufacture Keep out of direct sunlight This device should be disposed of separately from normal household waste so that components can be recycled Refer to User Manual before operating...

Страница 14: ...ew materials and used materials equivalent to new in performance and reliability MTTS may repair or replace MTTS products a with new or previously used products or parts equivalent to new in performan...

Страница 15: ...all software patches run MTTS diagnostics and utilities Implement temporary procedures or workarounds provided by MTTS while MTTS works on permanent solution Cooperate with MTTS in attempting to resol...

Страница 16: ...D SERVICES Co LTD House No 26 Alley 41 An Duong Vuong Street Tay Ho District Hanoi City Vietnam Tel 84 24 3766 6521 Fax 84 24 3718 8050 Email assistance mtts asia com www mtts asia com PROMETEO SRL Vi...

Отзывы: