Page 34
MTI Part Number 7001-0040
Revision 3.1 August 7, 2014
TOOLS Menu – Calibration Wizard Continued
1500CS Phase calibration
Next, disconnect the voltmeter, and
connect the 1500CS to the Phase
meter using two (2) BNC cables as
illustrated.
NOTE
The 1510A does not require phase
calibration. Skip this step for a
1510A.
Next, click in the Phase Reading
window and enter the reading of the
phase meter.
NOTE
If you do not have access to a
phase meter, verify that the value
–3.57 degrees is in the reading
window, and click the NEXT
button.
Next, click in the Phase Reading window
and enter the reading of the phase
meter.
NOTE
If you do not have access to a
phase meter, verify that the value
–4.65 degrees is in the reading
window, and click the NEXT
button.