Page 32
MTI Part Number 7001-0040
Revision 3.1 August 7, 2014
TOOLS Menu – Calibration Wizard Continued
Channel B calibration
Next, connect the BNC cable to the
1500CS/1510A Channel B output
connector and the input to the
voltmeter.
Set the voltmeter for AC RMS
readings.
Channel B Low range AC calibration
Next, click in the Voltage Reading
window and enter the reading of the
voltmeter. Verify that the correct
units (Pk, Pk-Pk, or RMS) are selected
for the type of voltmeter you are
using.
This screen is repeated for the three
low range voltages (14 mv, 35 mv,
and 56 mv) used.
Channel B Mid range AC calibration
Next, click in the Voltage Reading
window and enter the reading of the
voltmeter. Verify that the correct
units (Pk, Pk-Pk, or RMS) are selected
for the type of voltmeter you are
using.
This screen is repeated
for the three mid range
voltages (140 mv, 350
mv, and 560 mv) used.