background image

A.

 

Detector

 

 

If the zero or span adjustments cannot be made within the range of their respective 

potentiometers the detector needs to be replaced. To replace detector(s): 

 

1.

 

Remove the #10-32 thumb screws at right side of housing and remove detector’s 

latched holder. Pull holder out as far as wiring will allow. 

 

2.

 

Pull wire connectors off of tangs on the detector noting the color coding. Note: 

Detectors will pull out of holder with fingers.

 

3.

 

Discard old detector and attach  wires to new detector as noted in wiring code. If 

new CO detector was shipped with a shorting clip between the sensing and 

reference terminals, remove and discard clip before connecting the detector.

 

4.

 

Allow at least one hour for stabilization then calibrate according to the instructions 

supplied with your calibration kit.

 

 

 

B.

 

Replacement Problems 

 

If problems occur in replacing the above items or in calibration, consult the factory @   

1-800-542-6646. 

 

C.

 

Other Components 

 

Other components are not field replaceable without unsoldering. If repair is required, 

the entire instrument should be returned to factory. 

 

 

 

 

 

 

80367-7/15 

 