6
6056 Rev B
3.0
SPECIFICATIONS AND OPERATING INFORMATION
3.1
Specifications
Classification Class
1
Dimensions: Power Box
10.00˝ tall x 6.50˝ wide x 3.00˝ deep
Heater Plate
14.25˝ tall x 5.00˝ wide x 1.125˝ deep
Heater Plate Capacity
1000 mL Industry standard IV fluid bags.
Mounting: Heater Plate
Power Box
Hook for I.V. Pole
I.V. Pole C-Clamp, Bed Rail Bracket or Wall
Mount Bracket
Construction
Powder Coat Finish Aluminum
Factory Preset Set Point Temperature
40°C (104°F)
3.2
Storage and Operating Conditions
3.3
Electrical Requirements
Storage Temperature
-40 to 80°C (-40 to 176°F)
Humidity Conditions
0 to 90% up to 40°C (104°F) non-condensing.
10 to 55% at 55°C (131°F) non-condensing.
Ambient Operating Temperature
-10 to 55°C (14 to 131°F)
Power Requirements
120 VAC
60 HZ
Fused 250
VAC
2.0 AMP Time Delay
Amperage Consumed
1.09 AMPS
Heater Wattage
53 watts
Thermal Cutout Temperature
41°C (106°F) Automatic reset