MSI 4300/5 Скачать руководство пользователя страница 19

M  E  A  S  U  R  E  M  E  N  T    S  Y  S  T  E  M  S    I  N  T  E  R  N  A  T  I  O  N  A  L

MSI-4300 Porta-Weigh Plus/5 Key  

•  

User Guide  

   Page 19

1) Enable Calibration by pushing the key behind the

seal port.

2) Use  the  SELECT  (NET/GROSS)  key  to  scroll  to

the “HI SETPT” (or “LO SETPT”) message. Push
ENTER (TEST).

3) The  message  display  reads  “HI  VALUE.”  Push

ENTER.

4) The most significant digit flashes. The display reads

“INPUT WT.” Push the SELECT key until the first
digit of the desired weight value is displayed. Push
ENTER. (The CLEAR (ZERO) key can be used to
change an entry if necessary). Note: To enter 100
kgs on a 10,000 kg scale, the first two digits should
be entered as zeros (00100).

5) The  next  most  significant  digit  flashes.  Use  the

SELECT key and the ENTER key to finish out the
entry. Note: With the IR remote, the value can be
entered directly.

6) After the entire Hi Set Point is entered the display

will step to “MODE.” Push ENTER. The first se-
lection  is  “DISABLED.”  This  mode  allows  the
user  to  program  a  setpoint  without  the  setpoint
actually  operating.  To  enable  the  setpoint,  push
SELECT  until  either  the  “GROSS”  mode  or  the
“NETGROSS” mode is selected. In the “GROSS”
mode  the  setpoint  will  operate  at  the  value  set
based on Gross Weight regardless of any tare value
being subtracted. If this is the desired setpoint op-
eration mode push ENTER. If the setpoint should
operate relative to a tared weight the “NETGROSS”
mode  should  be  used.  Use  the  SELECT  key  to
change from “GROSS” to “NETGROSS” and then
push ENTER. In this mode the setpoint will oper-
ate  at  the  value  which  represents  either  a  Gross
weight or a Net weight.

7) The  next  selection  is  the  Setpoint  “RESPONSE”

— what the scale should do when a setpoint value
is reached. Push ENTER. There are now three se-
lections:  The  “NORMAL”  response  turns  on  the
internal  MOSFET  which  is  connected  to  a  user-
supplied alert device such as a siren or strobe light.
Use  the  SELECT  key  to  change  the  response  to

INDIVIDUAL DIGIT FLASHES EVERY 

TIME SELECT KEY IS PUSHED.

Содержание 4300/5

Страница 1: ...User Guide Quality Industrial Weighing and Force Measurement Equipment Measurement Systems International PORTA WEIGH PLUS CRANE SCALES 5 KEY VERSION MSI4300 5...

Страница 2: ...M E A S U R E M E N T S Y S T E M S I N T E R N A T I O N A L Page 2 MSI 4300 Porta Weigh Plus 5 Key User Guide...

Страница 3: ...Value 13 Test Switch 13 Calibration and Setup 14 Switch Functions during Calibration Setup 14 Enabling Calibration and Setup 14 Calibrate Setup Menu 15 Standard Calibration 15 Fine Calibration 16 Cal...

Страница 4: ...d fully shock mounted and marine proof construction makes for a rugged and reliable overhead scale 6 Digit Weight Indication 8 Character Message Display Units lb kg t METRIC t International Motion Ico...

Страница 5: ...eric characters Also used for menu prompts Display illumination uses rugged long life light emmitting diode LED backlighting coupled with a transflective liquid crystal display providing optimum displ...

Страница 6: ...ACEMENT 1 Turn the MSI 4300 Off 2 With one hand hold the battery access door to prevent it from falling 3 Rotate the two cam lock knobs on the access door counter clockwise until the door latches dise...

Страница 7: ...vailable life Due to the small maintenance discharge imposed on the battery by the MSI 4300 electronics do not store the MSI 4300 with the battery inside the scale for extended periods longer than 7 1...

Страница 8: ...recommends obtaining a spare battery so one can always be on the charger BATTERIES Optional Alkaline Battery Power 8 Standard D cells Alkaline type is preferred but standard Carbon Zinc can also be us...

Страница 9: ...M E A S U R E M E N T S Y S T E M S I N T E R N A T I O N A L MSI 4300 Porta Weigh Plus 5 Key User Guide Page 9 This page intentionally left blank...

Страница 10: ...zero any weight on the scale which causes the scale to go into the NET mode When in NET mode pushing TARE will clear the TARE value and the scale will revert to the GROSS mode 5 TEST When pushed this...

Страница 11: ...all internal operations are checked and any nonconformance causes an error message to be displayed 4 UK LFT units only ZERO User is prompted to push the zero key 1 20 zero range If the weight is in i...

Страница 12: ...Rules for Use 1 Does not operate in the GROSS mode 2 NET GROSS will work even when the scale is in motion Action Not in NET mode Tare No Action display continues to read the Gross weight only In NET m...

Страница 13: ...ibration number Action To start the text sequence push TEST 1 All digits 7 segment and 16 segment count once from 0 to 9 All annunciators are tested 16 segment displays also go through 2 Battery condi...

Страница 14: ...n 2 ZERO Used to zero the scale during calibration or CLEAR during setups 3 NET GROSS Used to scroll through menu items Used in FINE CAL as UP 4 TARE IN OUT Used only in FINE CAL as DOWN 5 TEST Used a...

Страница 15: ...ST eight steps with numerics at 123456 and message reading CONTRAST EXIT Push ENTER TEST to quit setup HI SET PT HI VALUE DISABLED GROSS NETGROSS RESPONSE NORMAL BLANKLCD BLINKLCD EXIT LO SET PT LO VA...

Страница 16: ...2 Use the SELECT NET GROSS key to scroll to the FINE CAL message Push ENTER TEST 3 Pick up a test weight of at least 50 of capacity The weight is indicated on the numeric digits WARNING This function...

Страница 17: ...pushing the key behind the seal port 2 Use the SELECT NET GROSS key to scroll to the AZM message Push ENTER TEST 3 Use the SELECT NET GROSS key to scroll through the AZM options When the desired mode...

Страница 18: ...it off LIGHTING The light emitting diode LED backlight is normally controlled by a photocell that detects ambient light and determines if the backlight should be on or off AUTOMATIC mode Should the us...

Страница 19: ...Push ENTER The first se lection is DISABLED This mode allows the user to program a setpoint without the setpoint actually operating To enable the setpoint push SELECT until either the GROSS mode or th...

Страница 20: ...peration use SELECT until the message EXIT appears and push ENTER or push CLEAR no shift required This returns you to the Setup Menu Again push CLEAR to get out of the menu and back to standard scale...

Страница 21: ...immediately by the Cal Switch The Scale reads RESET 2 Use the SELECT NET GROSS key to scroll to UNITS Push ENTER TEST 3 Use the SELECT NET GROSS key to scroll to the desired unit of measure Push ENTE...

Страница 22: ...achieve the maximum operation range The shape of the remote is designed to help in aiming Aiming becomes less critical as the distance decreases The differences between front panel operation and remot...

Страница 23: ...While in NET mode push TARE followed by PRINT Caution Do not accidently push the Tare key while in Gross mode as this will clear the old tare and establish a new tare Numeric Entry The Tare memory val...

Страница 24: ...and the battery is fresh 7 Replace the front panel switch assembly by centering it over the open area and pressing evenly around the edges until it seals 8 Replace the three screws Make sure the scre...

Страница 25: ...he number entry otherwise the ID will remain as before As shipped any Measurement Systems International infrared remote will operate any Porta Weigh Plus It is possible by special order to encode the...

Страница 26: ...MORY CLEAR SELECT TEST T 0 ZERO NET GROSS TARE TOTAL POWER VIEW SHIFT DO NOT USE MULTIPLE ATTACHMENTS DO NOT ENTER SETUP PEAK HOLD TARE MEMORY CLEAR SELECT TEST T 0 ZERO NET GROSS TARE TOTAL POWER VIE...

Страница 27: ...r alterations made by others except those repairs or alterations made with its specific written consent nor shall the Company be liable for any damages or delays whether caused by defective workmanshi...

Отзывы: