ML41biological microscope user manual
22
Website
is not smooth
properly placed
Slice clips not securely
fastened
Fasten the slice clip
The image moves
noticeably
when
touching the stage
Stage is not fixed properly Fixed the stage
6.Maintenance
6.1 The instrument must be placed in a dry, cool, ventilated place, and be careful to avoid
moisture and heat.
6.2 Pay attention to cleaning the optical parts frequently. For dust, you can use a blower
ball or a soft brush to clean it, or gently wipe it with a clean gauze. If you want to remove
fingerprints and oil stains, please use a clean gauze or lens tissue dipped in a small amount
of pure ethanol-ether mixture (mixed in a ratio of 3:7) to gently wipe off.
6.3 Do not use organic solvents to wipe the non-optical surfaces of the microscope. Use
gauze dipped in clean water to wipe these surfaces. For stubborn stains, neutral detergents
can be used to clean them.
☆
Both ethanol and ether are extremely flammable. Be careful not to place these chemicals
near working light sources, open flames or possible sources of electrical sparks, such as
when electronic equipment is turned on and off, try to use these chemicals in a
well-ventilated room .
6.4 If the microscope is wet with liquid when working, please cut off the power
immediately and wipe it dry.
6.5 After using the instrument, cover it with a dust cover. If the microscope will not be
used for a long time, it is best to remove the eyepiece and objective lens and put them in a