2

A. Standard Specification

B. Component parts

Figure 01

Function description

1.Switch lever——Used to switch between observation mode of bright field and fluorescence. Push in to get

fluorescence observation, pull out to get bright field observation.

2.Brightness knob—— Used to continously adjust brightness of the fluorescent illumination.
3.Power——Used control electrical power.

C.Installation Guide

To match with different brand microscope, please make sure your microscope brand can connect with the illuminator.

Take Nikon E100 as sample to introduce how to install the LED epi-fluorescence illumination.

1.Unpack and take out appropriate accessories includes epi-fluorescence illumination body, power adapter,

screwdriver.Then remove the microscope observation tube. As shown in Figure 02.

Mode

Color

LED central

wavelength

Excitation

filter

Dichroic

mirror

Emission

filter

MF-UV-LED

UV

365nm

330-380nm

>400nm

>420nm

3.Swith lever

2.Brightness knob

1.Power

Содержание MF-UV-LED

Страница 1: ...User Manual MF LED Epi LED Fluorescent Illumination Guangzhou Micro shot Technology Co Ltd www m shot com Sales m sho com ...

Страница 2: ... this manual carefully before operating this unit MSHOT MF UV LED Epi LED fluorescent illumination is used for upgrading upright infinity optical microscope to epi fluorescent functional it has UV LED excitation without affecting the bright field observation LED as light source has a long working life energy efficient and easy to operate MF UV LED not only can match with MSHOT microscope also can ...

Страница 3: ...C Installation Guide To match with different brand microscope please make sure your microscope brand can connect with the illuminator Take Nikon E100 as sample to introduce how to install the LED epi fluorescence illumination 1 Unpack and take out appropriate accessories includes epi fluorescence illumination body power adapter screwdriver Then remove the microscope observation tube As shown in Fi...

Страница 4: ... in power adapter to the DC block at the back of LED light box Get it power on and start to enjoy Adjust steps please consider D Using guide D Using guide 1 Push in switch lever and block light rod put a white paper on the stage of microscope turn microscope objective to 10X 2 Find power button turn on power push in switch lever There will be a circular light spot appears on the white paper Observ...

Страница 5: ... of difficulty of the sample was observed generally brightness adjust to the extent of 80 is good to use also less than full load current can prolong the service life of LED lamp 2 When doing fluorescence observation block transmitted light illumination condenser lens with non reflective black small plate to prevent interference with reflective light Or you can adjust the position of the condenser...

Отзывы: