![MSA OptimAir MM PAPR Скачать руководство пользователя страница 10](http://html.mh-extra.com/html/msa/optimair-mm-papr/optimair-mm-papr_operating-manual_1846579010.webp)
(1)
Line up the key in the charger cable plug with the slot in the female connector on the
battery pack.
(2)
Push the plug into the female connector and turn the ring to secure it.
(3)
Plug the charger into a 115-120 V, 50, 60 Hz (standard AC) outlet. The charger can be used internationally with the
appropriate adapter.
CAUTION!
Voltages lower than 115V are not recommended. Voltages greater than 120V will damage the battery charger and battery pack.
Failure to follow this caution can result in battery or equipment damage.
(4)
Charge the battery pack until a full charge is indicated.
LED Signaling
Condition
LED Indication
No Battery
Yellow
Charge Initialization
Yellow
Rapid Charge
Orange
Charge Top-Off
Green/Yellow
Trickle Charge
Green
Charge Error
Orange/Green
The charger uses one LED. When the battery is charging, the LED will be orange. When the LED is green the battery is fully
charged. The battery can stay connected to the charger until it is needed. If the LED is yellow, the battery is either not
connected or not at the proper temperature. The charging will start when the battery pack reaches the optimum charging
temperature of 35°F to 100°F. If the LED is orange/green , there is a failure. Disconnect battery pack, and unplug charger. Plug
charger back into outlet and reconnect battery pack to reset charger. If the LED is orange/green during any part of the charge,
after it has been reset, remove the battery from service.