The Calibration Options menu allows the user to:
• modify the calibration cylinder settings
(CYLINDER SETUP)
• enable/disable calibration due date and set the number of days
(CAL DUE OPTIONS)
• enable/disable the option to show the last cal date at turn-ON
(LAST CAL DATE)
• enable/disable the option for password-protected calibration
(CAL PASSWORD).
1. Press the:
• ▼ button go to next page
• ▲ button to go previous page
•
button to enter setup.
Setting Calibration Cylinder
This option has a dialog similar to the span calibration dialog.
The display shows all active sensors.
1. Press the button to enter setup.
• The screen for the first calibration cylinder displays.
2. Press the:
• ▼ or ▲ button to change the value.
•
button to confirm the setup.
With this confirmation, the instrument automatically moves to the
next cylinder setting.
3. Repeat the sequence for changing the required settings for all
necessary gas values.
• After the last setting is performed, the instrument returns to the
Calibration Options menu.
3-7
Содержание altair 5
Страница 68: ...6 3 Instrument Replacement Parts 6 3 ...
Страница 71: ...7 Appendix Flow Charts 7 1 Start Up Sequence Power ON 7 1 ...
Страница 72: ...7 2 Normal Operation 7 2 ...
Страница 73: ...7 3 Options from Main Page 7 3 ...
Страница 74: ...7 4 Password Protected Options 7 4 ...
Страница 75: ...7 5 Menu Sequence from Main page 7 5 ...
Страница 76: ...7 6 ...
Страница 77: ...7 6 Calibration Sequence 7 7 ...
Страница 78: ...7 7 Calibration Options 7 8 ...
Страница 79: ...7 8 Alarm Options 7 9 ...
Страница 80: ...7 8 1 Sensor Alarm Options 7 10 ...
Страница 81: ...7 9 Instrument Options 7 11 ...
Страница 150: ...6 3 Pièces de rechange pour l appareil 6 3 ...
Страница 232: ...6 3 Piezas de repuesto 6 3 ...