11.1.2 47K-HT
Relative response factors of tested gases with reference to Propane.
These values are only valid for new sensors and, unless otherwise stated, refer to
an ambient temperature of 110 °C. In this case the displayed values may vary by
up to ±20% from the target gas concentration.
Sample gas
100% LEL
in vol%
Propane relative
response factor
Response
time [secs]
t
50
*)
Response
time [secs]
t
90
*)]
Propane 1.7
1.0
≤
6
≤
12
2-Butanon 1.8
1.09
≤
5
≤
12
Acetone 2.5 0.92
≤
4
≤
11
Ethanol 3.1 0.82
≤
4
≤
10
Ethyl acetate
2.2
1.11
≤
6
≤
12
Toluol 1.1 1.32
≤
6
≤
12
*) Response times are with a flow through adapter and a gas flow of 1 l/min