23
APPENDIX:
The check valve of the connecting piece is leak tested with the Leak Tester:
D1118 829. The wetted check valve is tested in the following procedure:
•
Fill a little bit of water into the connecting piece and shake off excessive water.
•
Fix the connecting piece vertically (check valve on bottom) and connect leak
tester with adapter D1118 023 to bottom side of connecting piece (see illustra-
tion).
•
Press the bellows of leak tester behind pressure gauge until 7 mbar pressure
are built-up.
•
Close cock and observe pressure drop
The check valve is tight, if the pressure drop in one minute is not more than 1 mbar.
Содержание SSR 30/100 TR
Страница 1: ...Trainingsger t SSR TR...
Страница 21: ...21 Bild 31 Beh lter verplomben Plombenband gem Ersatzteilliste...
Страница 25: ...Training Apparatus SSR TR...
Страница 39: ...15 Fig 19 Extend neck strap all the way Fig 20 Fix lead seal on lock lead seal set according to spare part list...