10
─
retrieved after deletion.
③
Save the curve: After the standard curve data detection is completed, click Save Curve
to save the curve.
④
Fitting curve: After the standard curve data detection is completed, you can view the
fitting curve to see the effect of standard curve establishment.
⑤
Sample detection: After selecting a standard sample, the fluorescence value of the
standard sample can be detected. As shown in Sample 01 above.
⑥
Delete data: Select a single detected fluorescence value and delete it to remove
unnecessary or mishandled fluorescence values.
Methods for establishing a standard curve:
a. Prepare a series of standard samples, 5-9 standard samples;
b. Create a new standard curve and enter the standard curve name;
c. Select the fluorescence detection channel according to the sample type;
d. Enter the standard sample concentration into the table in Figure 6;
e. Detect the fluorescence value of the standard sample separately, pay attention to the
corresponding concentration and fluorescence value;
f. After the detection is completed, click the fitted curve to view the curve effect;
g. Save the standard curve, and then call the standard curve detection position sample in
the detection interface of Figure 5.