SI-500 Mix Smart Operational Manual
Chapter 2 Specification
─
2
─
Chapter 2 Specifications
1. The normal operating condition:
Ambient temperature
:
:
:
:
4
°°°°
C
~
~
~
~
45
°°°°
C
The relative humidity:
≤
70%
Power supply
:
:
:
:
24VDC 2.0 A
2. The basic parameters and the function
3. Mixing load limitation
The maximum permitted total weight of the mixing load is limited.
1) Mix the DWP and the tube holders 0.5 ml, 1.5/2.0 ml and PCR 96 with max.
2000rpm.
2) It only mix at a load between 80-300g with a maximum mixing frequency of
2000rpm.
3) When the load is too much and even the speed is more than 2000rpm, it reduces
the mixing frequency automatically to 1400rpm. In the display the message Too Fast
appears.
Mixing frequency
300~3,000rpm
(
(
(
(
in 10rpm increments
)
)
)
)
Touch vortexing frequency
3,500rpm
Adjustable mixing time
1s~ 99h59min
Mixing and vortexing radius
1.5mm
(
(
(
(
3mm mixing stroke
)
)
)
)
Noise level
<50 db
Power consumption
40W
Weight
5.5 Kg
Dimension
(
(
(
(
LXWXH
)
)
)
)
232X174X148mm