3)Press the " RH/Temp Button " ( 3-7, Fig. 1 ) once,
the " Temp. indicator " ( 3-11, Fig. 1 ) will light, the
" Display " ( 3-1, Fig. 1 ) will show the Temp. value.
* Press the " RH/Temp Button " ( 3-7, Fig. 1 ) once again,
the " Display " will return the " Humidity value ", in the
same time the " %RH indicator " ( 3-10, Fig. 1 ) will light again.
4-3 1st layer setting procedures
LoLt
Low Limit
HILt
High Limit
Low Limit Value Setting
1)Press the " Set Button " ( 3-4, Fig. 1 ) once, the "
Display " will show " LoLt ", now the meter is ready
for the Humidity " Low Limit " value setting.
Press the " RH/Temp Button " ( 3-7, Fig. 1 ) once,
the " Temp indicator " ( 3-11, Fig. 1 ) will light, now
the meter is ready for the Temperature " Low Limit "
value setting.
Remark :
* Under " Display " show " LoLt ", if " % RH indicator "
( 3-10, Fig. 1 ) is lit, meter is ready for " Humidity
Low Limit " setting.
* Under " Display " show " LoLt ", if " Temp. indicator "
( 3-11, Fig. 1 ) is lit, meter is ready for
" Temperature Low Limit " valuesetting.
* The function of " Low Limit value " setting, refer to
page 10, Fig. 2.
6
Содержание PHT-3109
Страница 18: ...7 THE ADDRESS OF AFTER SERVICE CENTER 15...