background image

Troubleshooting Guide 

Caution:

 

Always disconnect the power cord before troubleshooting.

 

Trouble / error message 

Cause 

Remedy 

Motor is not working

 

1. Rotor type setting is not correct.
2. Foreign bodies block the motor.
3. Motor is defective.

Turn the main power switch off, and 
then turn it on again. 
1. Please check and re-set the rotor.
2. Sweep away foreign bodies.
3. Contract the distributor for repair.

Over heat

 

1. Rotor type setting is not correct.
2. Over time use.

3. Motor is defective.

Turn the main power switch off, and 
then turn it on again. 
1. Please check and re-set the rotor.
2. Stop using to let the centrifuge cool

down. And then restart it.

3. Contract the distributor for repair.

Over current

 

1. Rotor type setting is not correct.
2. Rotor is overloaded

3. Motor is defective.

Turn the main power switch off, and 
then turn it on again. 
1. Please check and re-set the rotor.
2. Reduce the loading on the rotor.
3. Contract the distributor for repair.

Over voltage

 

1. Rotor type setting is not correct.
2. Switching power is defective.

Turn the main power switch off, and 
then turn it on again. 
1. Please check and re-set the rotor.
2. Contract the distributor for repair.

1. When the motor is rotating, push

the button     .

1. Waiting for motor stop completely,

and then push the button     ,
and then close the lid to restart the
centrifuge.

Unbalance

 

1. Rotor is unbalanced.

2. The surface which the centrifuge is

on is not flat.

1. Open the lid and set the rotor to be

balanced.

2. Place the centrifuge on a flat

surface.

Lid open error

 

1. During operation, the lid is open.

1. Waiting for the motor stop

completely. And close the lid.

Speed error

 

1. Rotor type setting is not correct.

2. Motor, PC board, or switching

power is defective, so the speed
can’t reach to the setting value.

1. Open the lid. Check and reset the

rotor.

2. Contract the distributor for repair.

No display on the screen

 

1. Power is off.
2. Dead power output

3. Burnt fuse

1. Turn on the power.
2. Change to a different output.
3. Replace a new fuse.

Service Manual 

Содержание MDCEN-302-SMD

Страница 1: ...MDCEN 302 SMD Microprocessor Centrifuge User Manual PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 7 22...

Страница 2: ...unded outlet Turn on the main power switch on the right side of the centrifuge 2 Push the button and then the screen is illuminated Press the button to open the lid The icon is flash on the screen Ins...

Страница 3: ...ton again The tube adapter size value is flash Push the button or to choose the tube adapter size Push the button to enter the rotor value Rotor type Rotor type on the screen Tube adapter size AR 1512...

Страница 4: ...s no number beside the icon Note Another way to exit the memory mode Push the button to turn off the power and then push the button again to exit the memory mode 12 Under the conditions of power failu...

Страница 5: ...ing can be done with a damp cloth Avoid the use of solvents as they may damage the product housing rotor 2 Check if the rotor is in the center In case the rotor is slanted Stop using machine and conta...

Страница 6: ...the rotor 3 Contract the distributor for repair Over voltage 1 Rotor type setting is not correct 2 Switching power is defective Turn the main power switch off and then turn it on again 1 Please check...

Страница 7: ...s crashed 1 Contract the distributor for repair 2 Contract the distributor for repair 3 Turnoffthemainpowerswitch and turnonthepoweragain Noise is too loud 1 Rotor is unbalanced 2 Rotor type setting i...

Отзывы: