13

 

Figure 4.2.1 

Figure 4.2.2 

4.2.2.2    Start Refrigeration System at -80

 

Press  power  switch,  the  temperature  display  window  PV,  SP  will  shows 

(Figure4.2.3). The  PV  window  will  shows  the  actual  measured  temperature(Figure  4.2.4)  after 

1s, SV window will shows the previous set temperature (Figure 4.2.4). Users don’t need to set 

SP temperature. 

Figure 4.2. 3     

Figure 4.2. 4 

4.3

 

Start Refrigeration System 

4.3.1    Start Refrigeration System at -40

 

Press ‘

’ after set all the parameters, the compressor will start working after delay-time, the 

cooling system will working continuously. 

4.3.2    Start Refrigeration System at -80

 

Press ‘

’ after set  all the  parameters, the  indicate light on, cold1 light  blinking(on delay 

mode),  the  high  temperature  device  will  start  working  after  time-delay,  the  low  temperature 

device will start working when it reached set temperature and cold2 light on, the cooling system 

will working continuously after low temperature device start working. 

4.4 Drain 

Please  drain  liquid  regularly  according  to  recovered  material  amount.  Open  drain  valve 

Содержание COT Series

Страница 1: ...Operation Manual COLD TRAPS COT SERIES PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 9 20...

Страница 2: ...onfiguration 8 1 4 4 COT 5000L Configuration 8 1 5 CONTROL PANEL INSTRUCTION 9 1 5 1 COT 2000H COT 5000H Control Panel Instruction 9 1 5 2 COT 2000L COT 5000L Control Panel Instruction 9 2 UNPACKINGAN...

Страница 3: ...2 6 2 MAINTENANCE OF COOLING SYSTEM 15 6 3 MAINTENANCE REQUIREMENT 15 6 4 COMMON FAILURES AND TROUBLESHOOTING 16...

Страница 4: ...rasp of the detailed requirements in this manual Otherwise it must be used under the guidance of the person who has the related technology skills 1 2 2 Proper Use COT series cold traps are required to...

Страница 5: ...dangerous and will lead to death or serious injury Note It shows that the situation is very dangerous and will lead to injury Prohibit Get wet in the rain or splashing water Note Do not close to the r...

Страница 6: ...ll to avoid electric shock If found refrigerant leak please take ventilation measures to reduce refrigerant content in the air as soon as possible The recycling material should has no corrosive on SUS...

Страница 7: ...ife will be affected Do not contact refrigerant with food medicines tobacco etc to avoid personal injury 1 2 7 Safety Measures Please wear personal protective equipment such as protective glasses prot...

Страница 8: ...US304 knob 2 Sealing flange 3 Control panel 4 Wheel 5 Breaker 6 Power cord 7 Air outlet 8 Liquid outlet 9 Air inlet 1 4 2 COT 5000H Configuration Figure 1 4 2 1 SUS304 knob 2 Sealing flange 3 Breaker...

Страница 9: ...aling flange 3 Control panel 4 Wheel 5 Breaker 6 Power cord 8 Air outlet 9 Liquid outlet 7 Air inlet 1 4 4 COT 5000L Configuration Figure 1 4 4 1 SUS304 knob 2 Sealing flange 3 Control panel 4 Air out...

Страница 10: ...re display window 2 Up 3 Down 4 Set 5 Power Switch 6Cooling Switch 7Circulaion Switch disabled 1 5 2 COT 2000L COT 5000L Control Panel Instruction Figure 1 5 2 Control Panel 1 Measured value and Menu...

Страница 11: ...ace the device steady place more than 12h and then start debugging 3 Keep well ventilated the environment temperature should less than 35 Note Do not cover the air inlet outlet 4 Connect air outlet wi...

Страница 12: ...n working temperature range Safety protection Time delay over current over heat Dehumidification capacity Max 1kg water Max 1kg water Max 4kg water Max 4kg water Cooling capacity W 10 640 920 20 400 5...

Страница 13: ...electric shock or equipment damage if you don t connect the power supply as described in 4 1 4 2 Power on 4 2 1Close Breaker Set the breaker which located on the right side of the instrument to then...

Страница 14: ...tem at 40 Press after set all the parameters the compressor will start working after delay time the cooling system will working continuously 4 3 2 Start Refrigeration System at 80 Press after set all...

Страница 15: ...set before leaving the factory users do not change it 6 Maintenance and Management Ensure to make the equipment running normally to increase its service life users should do daily maintenance and mana...

Страница 16: ...up the cooling system condenser regularly Please follow the steps below 1 Turn off the power supply 2 Take down the screws then pull the side panel down 3 Clean up the dust on the condenser fin Sugge...

Страница 17: ...ensor breakdown Replace sensor 4 Temperature lose control Voltage fluctuation range large Restart the device after voltage stabilization Temperature controller connection failure Stop using it and con...

Страница 18: ...airflow or airflow abnormal The fan has not started Stop using it and contact us or professional maintenance personnel The fan has started The fan was stuck Blade deformation Our company reserves the...