background image

 

 

Assembly 

 

 
 
 
All the steps described for the assembly of the microscope must be undertaken with extreme care, 
and without forcing the placement of the distinct parts and elements of the microscope.  
 

A.  Condenser (10): Turn the coarse 

focus knob (19) until the stage (8) 
reaches its highest position. Turn 
the focussing knob on the 
condenser (Fig.1) until the 
condenser holder reaches its 
lowest position. Unscrew the 
knurled locking screw on the 
condenser, and insert the 
condenser in the mounting ring so 
that the aperture of the diaphragm 
(9) faces the front of the 
microscope and so can be 
adjusted easily. Once firmly 
placed in the mounting ring, firmly 
tighten the locking screw, and turn 
the condenser focusing knob until 
in its highest position. 

 
 
 

Condenser

set screw

 
 
 
 
Condenser 
focusing 
knob 

Fig. 1 

-4- 

Содержание B1-220ASC

Страница 1: ...PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC VER 02 11 10 INSTRUCTION MANUAL FOR B1 220ASC Biological Microscope...

Страница 2: ...echanism 7 Slide holder 8 Stage 10 Condenser 9 Diaphragm 15 Head 17 Arm 14 Eyepiece tube with diopter correctors 18 Knurled locking screws for securing specimen holder 19 Coarse Focus Knob 20 Fine Foc...

Страница 3: ...the series Unpacking All components of the microscope have been carefully packed to make sure they reach you in perfect condition We recommend that you do not discard any packing containers in case yo...

Страница 4: ...the focussing knob on the condenser Fig 1 until the condenser holder reaches its lowest position Unscrew the knurled locking screw on the condenser and insert the condenser in the mounting ring so tha...

Страница 5: ...microscope Firmly re tighten the knurled locking screw E Eyepieces 1 Remove the protective covers on the eyepiece tubes 14 Gently place the eyepieces in the tubes without touching the lens surface B1...

Страница 6: ...lease the moveable finger until the specimen is supported in place 4 Make sure that the specimen is in the optical path by adjusting the stage using the knobs controlling the X Y movements 21 5 Lookin...

Страница 7: ...must ONLY be used with the 100X objective which is the only objective prepared for it If any other objective comes into contact with immersion oil it must be cleaned immediately G Critical illuminatio...

Страница 8: ...otective cap that covers the port can be removed Fig 4 Adapter tube Light controlling rod Knurled screw Fig 3 Photo adapter Video adapter 3 Insert the adapter tube with the camera mounted on the verti...

Страница 9: ...OCEDURE NOT COVERED IN THIS INSTRUCTION MANUAL A Optical maintenance 1 Do not attempt to disassemble any lens component 2 Before cleaning any lens remove surface dust with a fine brush especially for...

Страница 10: ...s Fig 5 g Close hinged door and tighten locking screw 2 Changing the fuse a Using a flat screwdriver gently press on the slot located on the fuse cap Fig 1 and make a turn in the direction indicated b...

Страница 11: ...t of the lens should it come into contact with the slide An additional measure of security consists of an adjustable stopper on the ascending movement of the stage The stopper comes pre adjusted by th...

Страница 12: ...by a qualified technician IMAGE QUALITY PROBLEM CAUSE SOLUTI N No image Nosepiece not positioned properly Image too bright Turn until clicks into place Reduce the intensity of the light Poor resoluti...

Страница 13: ...sion Warrantee All MRC microscopes are warranted against any manufacturing defect for a 5 year period Damage occurring by any unauthorised repair work or occurring through misuse or modification of th...