Troubleshooting 
 
ELECTRICAL 
 
PROBLEM CAUSE 

SOLUTIÓN 

Bulb does not work. 

Plug outlet does not work 
 
Cable not connected 
Bulb burned out. 
Fuse blown. 
Wrong bulb. 

Repair by a qualified specialised 
technician. 
Connect cable. 
Replace bulb. 
Replace fuse. 
Replace by the correct bulb. 

Bulb burns out in short 
time. 

Voltage too high. 

Reduce light intensity to a minimum 
before turning the microscope on or 
off.  

Bulb burns out immediately. Wrong bulb. 

Replace with the correct bulb. 

Bulb flickers. 

The bulb is not correctly inserted into 
the socket. 
Bulb about to burn out.  
Fuse holder not locked into proper 
position. 
Loose connection at plug outlet. 

Insert correctly. 
 
Replace bulb. 
Close correctly. 
Repair by a qualified specialised 
technician. 

Fuse blows in short time. 

Wrong fuse. 

Replace with the appropriate fuse. 

Fuse blows immediately. 

Short circuit 

Repair by a qualified technician. 

 
IMAGE QUALITY 
 

PROBLEM CAUSE 

SOLUTIÓN 

No image. 

Nosepiece not positioned properly. 
Image too bright. 

Turn until clicks into place.  
Reduce the intensity of the light. 

Poor resolution. 

Dirty objective. 
Dirty eyepiece. 
Slide upside down. 
 
Wrong cover slip used with slide.  
Light too bright. 
 
Dirty condenser. 

Clean objective. 
Clean eyepiece. 
Replace the slide with the cover slip 
facing upwards. 
Use 0.17mm thick cover slips. 
Reduce light intensity or adjust the 
diaphragm aperture. 
Clean condenser. 

Spots in field of view 

Dirty eyepiece. 
Dirty slide. 
Dirty condenser. 

Clean eyepiece. 
Clean preparation. 
Clean condenser. 

Uneven illumination of field. Nosepiece not positioned properly. 

Diaphragm aperture not sufficiently 
open. 

Turn until it clicks into place. 
Adjust appropriately. 

 
MECHANICAL 
 

PROBLEM CAUSE 

SOLUTIÓN 

It does not stay in focus. 

The stage is sliding down on its own.   Adjust  the  tension of the coarse focus 

knob.  

It does not focus.  

The stopper on the ascending 
movement of the stage needs 
adjusting. 

Readjust the stopper. 

 
 
 
 

 

-12- 

Содержание B1-220ASC

Страница 1: ...PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC VER 02 11 10 INSTRUCTION MANUAL FOR B1 220ASC Biological Microscope...

Страница 2: ...echanism 7 Slide holder 8 Stage 10 Condenser 9 Diaphragm 15 Head 17 Arm 14 Eyepiece tube with diopter correctors 18 Knurled locking screws for securing specimen holder 19 Coarse Focus Knob 20 Fine Foc...

Страница 3: ...the series Unpacking All components of the microscope have been carefully packed to make sure they reach you in perfect condition We recommend that you do not discard any packing containers in case yo...

Страница 4: ...the focussing knob on the condenser Fig 1 until the condenser holder reaches its lowest position Unscrew the knurled locking screw on the condenser and insert the condenser in the mounting ring so tha...

Страница 5: ...microscope Firmly re tighten the knurled locking screw E Eyepieces 1 Remove the protective covers on the eyepiece tubes 14 Gently place the eyepieces in the tubes without touching the lens surface B1...

Страница 6: ...lease the moveable finger until the specimen is supported in place 4 Make sure that the specimen is in the optical path by adjusting the stage using the knobs controlling the X Y movements 21 5 Lookin...

Страница 7: ...must ONLY be used with the 100X objective which is the only objective prepared for it If any other objective comes into contact with immersion oil it must be cleaned immediately G Critical illuminatio...

Страница 8: ...otective cap that covers the port can be removed Fig 4 Adapter tube Light controlling rod Knurled screw Fig 3 Photo adapter Video adapter 3 Insert the adapter tube with the camera mounted on the verti...

Страница 9: ...OCEDURE NOT COVERED IN THIS INSTRUCTION MANUAL A Optical maintenance 1 Do not attempt to disassemble any lens component 2 Before cleaning any lens remove surface dust with a fine brush especially for...

Страница 10: ...s Fig 5 g Close hinged door and tighten locking screw 2 Changing the fuse a Using a flat screwdriver gently press on the slot located on the fuse cap Fig 1 and make a turn in the direction indicated b...

Страница 11: ...t of the lens should it come into contact with the slide An additional measure of security consists of an adjustable stopper on the ascending movement of the stage The stopper comes pre adjusted by th...

Страница 12: ...by a qualified technician IMAGE QUALITY PROBLEM CAUSE SOLUTI N No image Nosepiece not positioned properly Image too bright Turn until clicks into place Reduce the intensity of the light Poor resoluti...

Страница 13: ...sion Warrantee All MRC microscopes are warranted against any manufacturing defect for a 5 year period Damage occurring by any unauthorised repair work or occurring through misuse or modification of th...

Отзывы: