MRC 1425 Скачать руководство пользователя страница 15

 

15

3/  check calibration – using independent thermometer, 
follow instructions in section 6.5  

Unit will not heat up at all 

 

 

1/  verify that controller is asking for heat by looking for 
Heating light – if pilot light is not on continuously during 
initial start up, there is a problem with the controller  
2/  check amperage – amperage should be virtually at 
maximum rated (data plate) amperage  
3/  do all controller functions work? 
4/  is the Controller set high enough? – for diagnostics, 
should be fully clockwise with the Over Temperature light 
never on 
5/  has the fuse blown? Check fuse at inlet. 
6/  Units will need at least some vacuum in chamber to keep 
unit air tight - verify with Vacuum Gauge is above “0” 

Indicated chamber temperature 
unstable 

 

 

1/  ± 1. may be normal  
2/  is ambient room temperature radically changing – either 
door opening or room airflow from heaters or air 
conditioning ? – stabilize ambient conditions  
3/  sensor miss-located, damaged or wires may be 
damaged  - check mounts for Main and Controller  sensors, 
then trace wires or tubing between sensors and controls 
4/  calibration sensitivity – call Customer Service 
5/  Controller set too low – be sure that it is more than 5 
degrees over desired set point; check if Over Temperature 
light is on continuously; turn controller knob completely 
clockwise to see if problem solved then follow instructions in 
section 6.6 for correct setting   
6/  electrical noise – remove nearby sources of RFI 
including motors, arcing relays or radio transmitters 
7/  bad connection on temperature sensor or faulty sensor – 
check connectors for continuity and mechanical soundness 
while watching display for erratic behavior; check sensor 
and wiring for mechanical damage 
8/  bad connections or faulty solid state relay – check 
connectors for mechanical soundness and look for corrosion 
around terminals or signs of arcing or other visible 
deterioration  

Will not maintain set point 

 

 

1/  assure that set point is at least 5 degrees over ambient 
room temperature  
2/  see if ambient room temperature is fluctuating 

Display and Reference 
thermometer don’t match 

 

 

1/  calibration error – see section 6.6 
2/  Main temperature sensor failure – evaluate if Heating 
light is operating correctly  
3/  Main controller failure – evaluate if Heating light is 

Содержание 1425

Страница 1: ...INSTRUCTION MANUAL FOR Vacuum Oven 1425 1445 1465 PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC VER 02 10 10...

Страница 2: ...N 8 0 MAINTENANCE SECTION 9 0 TROUBLESHOOTING SECTION 10 0 PARTS LIST UNIT SPECIFICATIONS SCHEMATICS These units are general purpose vacuum ovens for professional industrial or educational use where t...

Страница 3: ...1 2 Inspect for concealed loss or damage on the unit itself both interior and exterior If any the carrier will arrange for official inspection to substantiate your claim Save the shipping crate until...

Страница 4: ...should help in identifying the use and function of the available user adjustable components 2 1 Indicates AC Power On 2 2 Indicates Manual Adjust 2 3 Indicates Temperature 2 4 Indicates Over Temperatu...

Страница 5: ...affect performance such as heat from radiators ovens autoclaves etc Avoid direct sun fast moving air currents heating cooling ducts and high traffic areas To ensure air circulation allow a minimum of...

Страница 6: ...ings and cracks and get on electrical parts or that may contain solvents that will harm the coatings Periodic cleaning is strongly recommended WARNING Never clean the unit with alcohol or flammable cl...

Страница 7: ...ounded adapter plug 4 4 Disconnect the unit from the electrical power source before attempting any repairs or component replacements 4 5 If a mercury thermometer is used and breakage should occur make...

Страница 8: ...to 10 and is adjustable across this scale If the set point is exceeded the Circuit Breaker Fuse will limit the rise to approximately 10 C above the set point selected 5 5 OVER TEMPERATURE LIGHT This...

Страница 9: ...9 Figure 2 1425 1445 1465...

Страница 10: ...Once a good vacuum seal is accomplished the door will hold itself shut and sealed until the chamber is returned to atmospheric pressure 6 4 Watch the VACUUM GAUGE and when the required vacuum is obtai...

Страница 11: ...maintain set point 7 5 Calibrating the Main Temperature It is recommended that calibration is done once the unit is installed in its working environment and has been stabile at set point for several h...

Страница 12: ...en to operate two hours or until it has started cycling normally before you begin using it NOTE Slight vapor or smoke may occur in the initial heat up This is the dissipation of protective coatings th...

Страница 13: ...rfaces DO NOT USE spray cleaners that might leak through openings and cracks and get on electrical parts or that may contain solvents that will harm the coatings WARNING Never clean the unit with alco...

Страница 14: ...e see if Heating light is on compare current draw to data plate 6 Main controller failure confirm with front panel lights that controller is calling for heat 7 Controller failure confirm with front pa...

Страница 15: ...ubing between sensors and controls 4 calibration sensitivity call Customer Service 5 Controller set too low be sure that it is more than 5 degrees over desired set point check if Over Temperature ligh...

Страница 16: ...of square Oven won t hold vacuum 1 check door gasket for alignment and damage wear or lack of compliance 2 assure all vent and feed valves are closed tightly 3 assure tight connections to pump OTHER c...

Страница 17: ...e definition of cleaning technique and maintenance schedule If this product should require service contact your GLstributor If return of the product be necessary an authorization number must be obtain...

Страница 18: ...EMI Filter 10 AMP N A 2800502 Fuse 10 AMP 3300502 N A Fuse 6 3 AMP 250V N A 103555 Fuse Holder 3300501 N A 1425 Inner Door Gasket 1445 1465 3450508 3450509 3450510 3450508 3450509 3450510 Outer Door G...

Страница 19: ...9 x 12 x 9 1445 20 5 x 28 2 x 26 25 12 x 20 x 12 1465 18 x 24 x 18 Capacity Cubic Feet 1425 0 6 1445 1 7 1465 4 5 Temperature Range Sensitivity 1425 1445 1465 Amb 5 to 240 C 1 0 C MODEL VOLTS AMPS VA...

Страница 20: ...20 WIRE DIAGRAM...

Страница 21: ...21...

Отзывы: