SECTION 7 
OPERATION

 

 
 

7.1 Power 

Supply:

  Connect the service cord to a grounded outlet and switch unit to 

the ON position.  If supplied with a detachable cord set, plug the female end into 
the inlet of the unit and the male end into the supply.  Assure that units requiring a 
fuse have a fuse installed.  This fuse may be at the inlet or part of the cord set male 
plug. 

 

7.2

 

Turn the Circuit Breaker/Fuse control to its maximum position (clockwise). 

 

7.3

 

Place a certified reference thermometer inside the chamber where it can be easily 
viewed through the window.  Vacuum down the chamber as described in Section 
5.0 

 

7.4 Setting 

Main 

Temperature:

 To enter set point mode on the control, press either 

the UP or DOWN arrow pad one time.  The digital display will start to blink from 
bright to dim. While blinking, the digital display is showing the set point. To change 
the set point, use the Up or Down arrow pads. If the arrow pads are not pressed for 
five (5) seconds, the display will stop blinking and will read the temperature of the 
unit.  The Heating Light Cycle light will come on indicating that the oven is heating.  
Allow the unit several hours to reach and maintain set point 

 

7.5 

Calibrating the Main Temperature:

  It is recommended that calibration is done 

once the unit is installed in its working environment and has been stabile at set 
point for several hours.  Compare the reading of the reference thermometer with 
the digital display.  If there is an unacceptable difference, put the display into 
calibrate mode by pressing both the UP and DOWN buttons at the same time until 
the decimal points start to blink.  While blinking, the display can be changed to 
match the reference thermometer by pushing the UP or DOWN buttons until the 
display reads the correct value.  If no buttons are pushed for five (5) seconds the 
display will revert back to reading the temperature in the chamber.  Allow for the 
oven temperature to stabilize again and recalibrate if necessary. 

 

7.6 

Setting the Circuit Breaker/Fuse:

 Once the Main Temperature is stabile at set 

point, set the Circuit Breaker/Fuse control.  Previously set to its maximum position, 
turn the control knob counterclockwise just until the OVER TEMPERATURE LIGHT 
comes on.  Next turn the control knob clockwise just until the light goes off.  Then 
turn the control knob clockwise two (2) minor scale divisions on the dial past the 
point where the light went out.  The Circuit Breaker/Fuse control will now be set 
approximately 10

°

C above the Main temperature set point. 

 

Содержание 1425

Страница 1: ...INSTRUCTION MANUAL FOR Vacuum Oven 1425 1445 1465 PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC VER 02 10 10...

Страница 2: ...N 8 0 MAINTENANCE SECTION 9 0 TROUBLESHOOTING SECTION 10 0 PARTS LIST UNIT SPECIFICATIONS SCHEMATICS These units are general purpose vacuum ovens for professional industrial or educational use where t...

Страница 3: ...1 2 Inspect for concealed loss or damage on the unit itself both interior and exterior If any the carrier will arrange for official inspection to substantiate your claim Save the shipping crate until...

Страница 4: ...should help in identifying the use and function of the available user adjustable components 2 1 Indicates AC Power On 2 2 Indicates Manual Adjust 2 3 Indicates Temperature 2 4 Indicates Over Temperatu...

Страница 5: ...affect performance such as heat from radiators ovens autoclaves etc Avoid direct sun fast moving air currents heating cooling ducts and high traffic areas To ensure air circulation allow a minimum of...

Страница 6: ...ings and cracks and get on electrical parts or that may contain solvents that will harm the coatings Periodic cleaning is strongly recommended WARNING Never clean the unit with alcohol or flammable cl...

Страница 7: ...ounded adapter plug 4 4 Disconnect the unit from the electrical power source before attempting any repairs or component replacements 4 5 If a mercury thermometer is used and breakage should occur make...

Страница 8: ...to 10 and is adjustable across this scale If the set point is exceeded the Circuit Breaker Fuse will limit the rise to approximately 10 C above the set point selected 5 5 OVER TEMPERATURE LIGHT This...

Страница 9: ...9 Figure 2 1425 1445 1465...

Страница 10: ...Once a good vacuum seal is accomplished the door will hold itself shut and sealed until the chamber is returned to atmospheric pressure 6 4 Watch the VACUUM GAUGE and when the required vacuum is obtai...

Страница 11: ...maintain set point 7 5 Calibrating the Main Temperature It is recommended that calibration is done once the unit is installed in its working environment and has been stabile at set point for several h...

Страница 12: ...en to operate two hours or until it has started cycling normally before you begin using it NOTE Slight vapor or smoke may occur in the initial heat up This is the dissipation of protective coatings th...

Страница 13: ...rfaces DO NOT USE spray cleaners that might leak through openings and cracks and get on electrical parts or that may contain solvents that will harm the coatings WARNING Never clean the unit with alco...

Страница 14: ...e see if Heating light is on compare current draw to data plate 6 Main controller failure confirm with front panel lights that controller is calling for heat 7 Controller failure confirm with front pa...

Страница 15: ...ubing between sensors and controls 4 calibration sensitivity call Customer Service 5 Controller set too low be sure that it is more than 5 degrees over desired set point check if Over Temperature ligh...

Страница 16: ...of square Oven won t hold vacuum 1 check door gasket for alignment and damage wear or lack of compliance 2 assure all vent and feed valves are closed tightly 3 assure tight connections to pump OTHER c...

Страница 17: ...e definition of cleaning technique and maintenance schedule If this product should require service contact your GLstributor If return of the product be necessary an authorization number must be obtain...

Страница 18: ...EMI Filter 10 AMP N A 2800502 Fuse 10 AMP 3300502 N A Fuse 6 3 AMP 250V N A 103555 Fuse Holder 3300501 N A 1425 Inner Door Gasket 1445 1465 3450508 3450509 3450510 3450508 3450509 3450510 Outer Door G...

Страница 19: ...9 x 12 x 9 1445 20 5 x 28 2 x 26 25 12 x 20 x 12 1465 18 x 24 x 18 Capacity Cubic Feet 1425 0 6 1445 1 7 1465 4 5 Temperature Range Sensitivity 1425 1445 1465 Amb 5 to 240 C 1 0 C MODEL VOLTS AMPS VA...

Страница 20: ...20 WIRE DIAGRAM...

Страница 21: ...21...

Отзывы: