VACUUM OPERATION 

 

 
 
IT IS IMPORTANT TO USE VACUUM TUBING FOR ALL 
VACUUM HOOKUPS. OTHER TYPES OF TUBING MAY 
COLLAPSE AND PREVENT COMPLETE EVACUATION.

 

 

6.2

 

A pump with a pumping capacity four times greater than the chamber 
volume is advisable. For example a 1410 has a chamber volume of one 
(1) cubic foot so a pump with a pumping with a pumping capacity of 
four (4) cubic feet per minute is recommended. When working below 
1mm, a diffusion type pump will be needed. See unit specifications  
for chamber capacities. 

 

6.3

 

Vacuum 

: To apply vacuum to the chamber attach the hose from the 

vacuum pump to the larger 3/8” hose connection on top of the oven. 
Close the VENT valve and open the VACUUM valve. Latch the door 
shut and start the vacuum pump. Be certain the vacuum valve is open 
and the VENT valve is closed. This action will hold the door shut and 
against the gasket until the pump creates a vacuum in the chamber. 
Once a good vacuum seal is accomplished, the door will hold itself 
shut and sealed until the chamber is returned to atmospheric pressure. 

 

6.4

 

Watch the VACUUM GAUGE and when the required vacuum is 
obtained, close the VACUUM valve and turn the pump off. The 
VACUUM GAUGE is calibrated from zero to 30 inches of Hg 
(762mm of Hg) with zero representing atmospheric pressure. The 
oven can be evacuated to pressures as low as 10 microns. 

 

6.5

 

Vacuum Release 

: To return the chamber to atmospheric pressure, 

open the VENT valve very slowly and allow the chamber to re- 
pressurize. The speed of pressurizing can be controlled by how much 
the valve is opened. 

Содержание 1407 DIG

Страница 1: ...Instruction Manual For Digital Vacuum Oven 1407 1408 DIG PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 11 18...

Страница 2: ...ture controller marked SET TEMPERATURE Its dial is marked from 0 to 10 and is adjustable across this scale A clockwise adjustment raises the temperature 4 4 VACUUM This adjustment valve located on the...

Страница 3: ...Front View Rear View...

Страница 4: ...emove the ground prong from the power cord or use an ungrounded 2 prong adapter plug 5 4 Disconnect the unit from the electrical power source before attempting to make any repairs or component replace...

Страница 5: ...ACUUM valve Latch the door shut and start the vacuum pump Be certain the vacuum valve is open and the VENT valve is closed This action will hold the door shut and against the gasket until the pump cre...

Страница 6: ...lower button to rise or lower the temperature setpoint SV The new setpoint is entered when the button is released and indicated by a brief flash of the display The actual temperature measured at the p...

Страница 7: ...llow the unit to stabilize for several hours Temperature stability is obtained when the HEATING light circulates on and off to maintain set point and the temperature value in the chamber remains consi...

Страница 8: ...se a disinfectant that is suitable to your Application DO NOT use chlorine based bleaches or abrasives as this will damage stainless steel surfaces B When washing the gasket handle carefully so as not...

Отзывы: