PRECAUTIONS

 

 

THIS IS NOT AN EXPLOSION PROOF OVEN

 

 

5.1

 

Do not place or use explosive, combustible, or flammable materials in 
the oven. 

 

5.2

 

Do not use sealed containers in the oven chamber. 

 

5.3

 

Do not cut or remove the ground prong from the power cord or use an 
ungrounded 2- prong adapter plug. 

 

5.4

 

Disconnect the unit from the electrical power source before 
attempting to make any repairs or component replacements. 

 

5.6 

THIS OVEN IS NOT SUITABLE FOR USE IN CLASS 1,2 OR 3 
LOCATIONS AS DEFINED BY THE NATIONAL 
ELECTRICAL CODE NFPA 70.

 

 

5.7 

This oven is not intended, nor can it be used, as a patient connected 
device. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание 1407 DIG

Страница 1: ...Instruction Manual For Digital Vacuum Oven 1407 1408 DIG PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 11 18...

Страница 2: ...ture controller marked SET TEMPERATURE Its dial is marked from 0 to 10 and is adjustable across this scale A clockwise adjustment raises the temperature 4 4 VACUUM This adjustment valve located on the...

Страница 3: ...Front View Rear View...

Страница 4: ...emove the ground prong from the power cord or use an ungrounded 2 prong adapter plug 5 4 Disconnect the unit from the electrical power source before attempting to make any repairs or component replace...

Страница 5: ...ACUUM valve Latch the door shut and start the vacuum pump Be certain the vacuum valve is open and the VENT valve is closed This action will hold the door shut and against the gasket until the pump cre...

Страница 6: ...lower button to rise or lower the temperature setpoint SV The new setpoint is entered when the button is released and indicated by a brief flash of the display The actual temperature measured at the p...

Страница 7: ...llow the unit to stabilize for several hours Temperature stability is obtained when the HEATING light circulates on and off to maintain set point and the temperature value in the chamber remains consi...

Страница 8: ...se a disinfectant that is suitable to your Application DO NOT use chlorine based bleaches or abrasives as this will damage stainless steel surfaces B When washing the gasket handle carefully so as not...

Отзывы: