MUNI User’s Guide
18
6.1.1 Configuration Details
These items can be set up using mPower Suite and are not accessible in the instrument
Configuration Mode or are easier to manage in mPower Suite.
•
Alarm Mode:
Auto Reset
means that the alarm signals stop as soon as the alarm condition
no longer exists.
Latched Mode
means the alarm continues until the user acknowledges it
by pressing the Left [+/OK] key. Under
Alarm Device
, the user can check or uncheck the
boxes to enable or disable any alarm signal.
WARNING!
Disabling all of the alarm devices prevents notification of hazardous gas concentrations and can
lead to serious injury or death!
•
Backlight:
Automatic
turns on automatically under dim light conditions and
Manual
requires the user initiation by pressing either the Left [+/OK] or Right [ /↓] keys.
•
Policy Check:
o
“
Must Bump/Cal
” locks the instrument from being used once the
Bump
or
Cal
Interval
is reached, until a Bump or Cal is performed.
o
“
Bump/Cal Due Lock
” allows overriding the Bump/Cal requirement by entering the
instrument password.
•
Bump
and
Calibration
Intervals
can be set in the far right panel, either under the
Summary tab if all sensors are the same, or individually under each sensor tab. The software
allows setting intervals between 1 and 365 days, but for all MUNI instruments calibration
intervals should not exceed 180 days. Notifications and Policy Check are disabled by setting
the intervals to zero.
mPower Suite Alarm, Span and Bump/Cal Interval Settings