18
Fault Finding
7 Fault Finding
FAULT
CHECK
Unexpected results
obtained from sample
Ensure that the minimess hose has been fully con-
nected at both the bottle sampler and the CMP.
Confirm that there is a free flow of fluid to the CMP,
by operating the flush valve and observing fluid pass-
ing to waste.
Check that the BS500 is reaching sufficient pressure,
2 to 2.5bar.
Bottle sampler not
reaching required
pressure.
Re-grease o-ring in the base of bottle sampler using a
suitable grease, silicone grease is recommended.