12
Fluid/Assembly Preparation and Analysis
5.2 Procedure
1. Connect minimess hose (600mm) to the contamination monitoring product.
2. Connect minimess hose (600mm) to the bottle sampler connection.
3. Connect waste fluid hose to contamination monitoring product (waste connector).
4. Insert waste fluid hose into the waste bottle provided.
5. Connect the power supply to bottle sampler, using power adaptor supplied.
6. Switch ON the contamination monitoring product.
7. Select Short Test Sampling option.
8. Enter test details, for more information refer to the contamination monitoring
product user guide.
9. Remove the top chamber by rotating anti clockwise, until it disengages and then
lift to remove.
10. Place a bottle with minimum 200ml of flushing fluid onto the base unit.
11. Replace the top chamber. The silver spot on the top chamber should be aligned
with the centre of the smallest groove. The top chamber can then be lowered and
the top should be lined up with the silver line between the two lines on the base.
12. Turn the selector handle to Pressure, towards the red disc P.
13. Turn the top valve anti-clockwise to the open position, towards the P.
14. Switch on the BS500.
15. The pressure inside the chamber will slowly increase and the visiwink will turn
red.
16. Once the BS500 stabillises, this takes about 30 seconds, press flush button on the
contamination monitoring product. This is approximately 30 seconds after the
unit is switched on.