Manufacturer’s declaration-electromagnetic immunity
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
eMOTION is intended for use in the electromagnetic environment (for home healthcare and professional healthcare)
specified below.
The customer or the user of the eMOTION should assure that it is used in such an environment.
Test
frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power
(W)
Distance
(m)
IMMUNITY
TEST
LEVEL
(V/m)
Compliance LEVEL
(V/m)
(for home healthcare and
professional healthcare
environment)
385
380
–390
TETRA 400
Pulse
modulation b)
18 Hz
1,8
0,3
27
27
450
430 –
470
GMRS 460,
FRS 460
FM c)
±5 kHz
deviation
1 kHz sine
2
0,3
28
28
710
704 –
787
LTE Band 13,
17
Pulse
modulation b)
217 Hz
0,2
0,3
9
9
745
780
810
800 –
960
GSM 800/900,
TETRA 800,
IDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation b)
18 Hz
2
0,3
28
28
870
930
1720
1700 –
1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse
modulation b)
217 Hz
2
0,3
28
28
1845
1970
2450
2400 –
2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation b)
217 Hz
2
0,3
28
28
5240
5100 –
5800
WLAN 802.11
a/n
Pulse
modulation b)
217 Hz
0,2
0,3
9
9
5500
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting
antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted
by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.
* Recommended separation distances between other equipment and this device – avoid stacking and locating
it near other electronic devices.
* If abnormal behavior is observed due to EM disturbances, please relocate the device accordingly.