Accuracy
Overload
Range
Resolution
Max. Reading
±(%R Digit)
Protection
600 mV
0.1 mV
599.9 mV
± (0.5% + 3)
6 V
1 mV
5.999 V
± (0.5% + 3)
1050V
60 V
10 mV
59.99 V
± (0.5% + 3)
DC/AC rms
600 V
100 mV
599.9 V
± (0.5% + 3)
1000 V
1V
1000 V
± (0.8% + 3)
DC VOLTAGE :
7.2 TECHNICAL SPECIFICATION
Input impedance: 10M ohms approx
ACCURACY :
Common Mode Rejection Ratio > 100dB at DC, 50 / 60Hz (1k Ohm
unbalanced).
Normal Mode Rejection Ratio > 50dB at DC, 50 / 60Hz.
Temp. Coefficient = (0.1% x specified) accuracy per degree Centigrade
0
refered to 25 C.
Accuracy is valid from 10% of range to 95% of the range.
0
0
Accuracy is specified as ± % of reading and digit is valid at 25 C ± 3 C,
55% ± 10% RH.
It is recommended to recalibrate the instrument once in a year.
Accuracy
Overload
Range
Resolution Max. Reading
±(%RDigit)
protection
60 mV
0.01 mV
59.99 mV
± (0.5% + 3)
440V
600 mV 0.1 mV
599.9 mV
± (0.5% + 3)
DC mV :
Input impedance: >1000M ohms approx
DC/AC rms
Relevant Standard(s) Specification(s) : EN61010-1: 2001,
: EN61326-1: 2005
Measurement Category
: CAT IV Reinforced Insulation
Directives for CE Certification
: LVD:2006/95/EC
: EMC:2004/104/EC
22