3.0 PRODUCT DESCRIPTION
The Pure-Vu System enables colon cleansing during colonoscopy using a standard or slim colonoscope with
a length of 1630mm – 1710mm and an outer diameter range of 11.7mm – 13.3mm (slim) and 12.8mm – 13.7mm
(standard). The Oversleeve, which fits over the colonoscope and is connected to an external Workstation,
generates fluid and gas to break up feces. The fecal matter & fluids are removed through the suction
channels of the Oversleeve into an external waste container.
The Pure-Vu System consists of the following main components:
• Disposable device which includes an Oversleeve and an Umbilical Section. The Oversleeve fits over
the colonoscope and the Umbilical section is connected to the external Workstation.
• Workstation that is reusable and supplies an irrigation mixture of water or saline and gas, and
evacuates fecal material & fluids.
• Loading Fixture that is reusable and aids in assembling the Oversleeve onto the colonoscope.
• Additional components include 60cc luer lock syringe/3-way stopcock assembly, sealing plug, and air
line tubing.
Below is a drawing showing the various components of the system and where they connect to each other.
1
Workstation
2
Irrigation Bag
3
Irrigation Line Clamp
4
Colonoscope
5
Oversleeve
6
In-Line Connector
7
Umbilical Section
8
Waste Container
9
Foot Pedal
10
Main Connector
11
Suction Tubing
Clamp
SYS TEST
LOW MED HIGH
ERROR
STATUS
DRAIN
1
2
4
3
5
6
8
7
9
10
11