MOTUS GI PURE-VU SYSTEM INSTRUCTIONS FOR USE
FIGURE 10:
Connect the two suction lines to the two waste containers (A) and ensure that at least one port, in
addition to the vacuum port (B), is open on each container. Ensure that the suction tubing clamp (C) is on the left of
the Workstation and then place the suction tubing (D) into the suction pump marked in green (E) between marked
green lines, then place the suction tubing (F) into the suction pump marked in red (G) between marked red lines,
then close both suction pump heads.
FIGURE 11:
Connect the Oversleeve to the Umbilical Section by the inline connector (A to B). Attach the inline
connector clip to the colonoscope Umbilical section (C to D).
4.2.3. Attaching the Umbilical Section to the Workstation
(cont...)
4.2.4. Attaching the Oversleeve to the Umbilical Section