background image

15

3.2.3 Centering the condenser (Models RED230 and 233)

• Fully open the field of view diaphragm and condenser aperture diaphragm. 
• Set the specimen on the stage with the cover glass facing up. 
• Bring the specimen image into focus, using the 10X objective.
• Close the field of view diaphragm to its minimum setting by means of the field diaphragm ring. 
• Turn the condenser focus knob to bring the field diaphragm image into focus on the specimen plane.
• Adjust the condenser centering screws so that the image of the field diaphragm appears at the centre  

of the field of view. At this time, stopping the field diaphragm image, just short of the maximum field of view,  
may be convenient for centering.

• Adjust and centre the field diaphragm so that it is just outside the field of view for each magnification change.

3.2.4 Use of field diaphragm (Models RED230 and 233)

Figure 7

• The field diaphragm determines the illuminated area on the specimen. Rotating the field diaphragm ring changes 

the size of the field diaphragm. For normal observation, the diaphragm is set slightly larger than the field of view. 
If a larger than required area is illuminated, extraneous light will enter the field of view. This will create a flare in 
the image and lower the contrast.

• The thickness of the glass slide must be 1.7mm.

1

Содержание RED 200 Series

Страница 1: ...Motic Incorporation Ltd Instructions Manual English RED200 SERIES...

Страница 2: ......

Страница 3: ...warranty 07 2 Instrument description 08 2 1 General views 08 2 2 Part names 09 2 3 Application 12 2 4 Instrument and its major features 12 3 First time use operation 13 3 1 First time use 13 3 2 Opera...

Страница 4: ......

Страница 5: ...g to the EN 61010 1 2001 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use 1 3 Unpacking transportation storage The original shipping container a foam box in a fi...

Страница 6: ...f the protection function If there is any electrical failure of the fuse system ground protection or transformer turn off and unplug the unit immediately Make sure the microscope is set aside so it wi...

Страница 7: ...found notify the manufacturer immediately Upon receipt of the Notification of Defect as described above the manufacture is responsible to solve the problem either by repairing the defective instrumen...

Страница 8: ...8 RED200 SERIES 2 1 General Views 2 INSTRUMENT DESCRIPTION RED230 233 Eyepiece Trinocular Head Binocular Head 10x 20 10x 18 RED220 223...

Страница 9: ...ry distance scale 8 Condenser fastening screw 3 Binocular head 9 Body tube lock screw 4 Quadruple Nosepiece 10 Arm and base a single piece 5 4X 10X 40X 100X objectives 11 Stage adjustment knob X axis...

Страница 10: ...ES 13 Condenser focus knob 17 Condenser aperture diaphragm adjustment handle 14 Coarse and fine focus knob 18 Collector RED230 RED233 15 Brightness control 16 Condenser RED230 RED233 RED230 RED233 13...

Страница 11: ...11 RED220 RED223 19 Condenser RED220 RED223 20 Collector RED220 RED223 19 20...

Страница 12: ...t Cord hanger at the back to accommodate power cable convenient and practical Coaxial coarse and fine focus adjustment with coarse focus tension control 75mm x 30mm mechanical stage with slide clips Q...

Страница 13: ...lugging in keep in mind that the working voltage of the biological microscope shall be the same as the supply voltage Figure 1 Turn on the power switch at the back of the base Figure 2 Note Make sure...

Страница 14: ...nation The RED200 Series biological microscope has been set already before delivery and can be adjusted as follows Put the specimen on the stage and fix it with the slide clips Note The thickness of t...

Страница 15: ...of the field of view At this time stopping the field diaphragm image just short of the maximum field of view may be convenient for centering Adjust and centre the field diaphragm so that it is just ou...

Страница 16: ...icroscope before making any modifications Figure 10a Figure 10b 1 2 3 3 2 Replace the objective Lower the stage all the way with the coarse focus knob Rotate the nosepiece to move the objective to be...

Страница 17: ...eads can be con nected to the photo port of the biological microscope using an adapter a 0 5X adapter is supplied Screw the adapter onto the camera Loosen the setscrew of the photo port and remove the...

Страница 18: ...o the slot Rotate the nosepiece to lock into the slot Condenser is not set properly Set the condenser properly Aperture iris diaphragm is not set accu rately Set the aperture iris diaphragm accurately...

Страница 19: ...rage Permissible environment temperature Permissible relative humidity 10 40 C Below 31 C max humidity is 80 at 40 C linearly decreases to 50 Operation Permissible environment temperature Permissible...

Страница 20: ...nser fixed K hler 140x135mm 75x30mm On the right On the right On the left of movable clip Used for 4X 100X objectives Binocular tube 30 20 Length of mechanical tube Maximum field number Hinge type int...

Страница 21: ...21...

Страница 22: ...ouse 311 Gloucester Road Causeway Bay Hong Kong Tel 852 2837 0888 Fax 852 2882 2792 Motic Spain S L U Spain Pol Ind Les Corts C Les Corts 12 08349 Cabrera de Mar Barcelona Spain Tel 34 93 756 6286 Fax...

Отзывы: