background image

13

3.1 First time use

Before installing and using the biological microscope, make sure to read the Safety Instructions (See Chapter 1) carefully.
When unpacking and handling, please do not touch the optical surfaces.

3. FIRST TIME USE & OPERATION

• After unpacking, place the biological microscope on a flat workbench and remove any foam padding or spacer 

used to prevent vibration during transportation. 

• Connect the cable at the base to the power supply. Before plugging in, keep in mind that the working voltage  

of the biological microscope shall be the same as the supply voltage. (Figure 1)

• Turn on the power switch at the back of the base. (Figure 2)  

Note: Make sure that the brightness control is in the minimum position before turning on or off the power switch.

• Rotate the brightness control to the desired illumination. (Figure 3)
• After use, turn the brightness control to the minimum position, and then turn off the power and put on the dust-

proof cover.

• The coarse focus tension (Figure 4) has been set at the factory, but can be readjusted as required  

(see figure 4, number 1).

Figure 2

Figure 3

Figure 4

Figure 1

1

2

1

3

2

Содержание RED 200 Series

Страница 1: ...Motic Incorporation Ltd Instructions Manual English RED200 SERIES...

Страница 2: ......

Страница 3: ...warranty 07 2 Instrument description 08 2 1 General views 08 2 2 Part names 09 2 3 Application 12 2 4 Instrument and its major features 12 3 First time use operation 13 3 1 First time use 13 3 2 Opera...

Страница 4: ......

Страница 5: ...g to the EN 61010 1 2001 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use 1 3 Unpacking transportation storage The original shipping container a foam box in a fi...

Страница 6: ...f the protection function If there is any electrical failure of the fuse system ground protection or transformer turn off and unplug the unit immediately Make sure the microscope is set aside so it wi...

Страница 7: ...found notify the manufacturer immediately Upon receipt of the Notification of Defect as described above the manufacture is responsible to solve the problem either by repairing the defective instrumen...

Страница 8: ...8 RED200 SERIES 2 1 General Views 2 INSTRUMENT DESCRIPTION RED230 233 Eyepiece Trinocular Head Binocular Head 10x 20 10x 18 RED220 223...

Страница 9: ...ry distance scale 8 Condenser fastening screw 3 Binocular head 9 Body tube lock screw 4 Quadruple Nosepiece 10 Arm and base a single piece 5 4X 10X 40X 100X objectives 11 Stage adjustment knob X axis...

Страница 10: ...ES 13 Condenser focus knob 17 Condenser aperture diaphragm adjustment handle 14 Coarse and fine focus knob 18 Collector RED230 RED233 15 Brightness control 16 Condenser RED230 RED233 RED230 RED233 13...

Страница 11: ...11 RED220 RED223 19 Condenser RED220 RED223 20 Collector RED220 RED223 19 20...

Страница 12: ...t Cord hanger at the back to accommodate power cable convenient and practical Coaxial coarse and fine focus adjustment with coarse focus tension control 75mm x 30mm mechanical stage with slide clips Q...

Страница 13: ...lugging in keep in mind that the working voltage of the biological microscope shall be the same as the supply voltage Figure 1 Turn on the power switch at the back of the base Figure 2 Note Make sure...

Страница 14: ...nation The RED200 Series biological microscope has been set already before delivery and can be adjusted as follows Put the specimen on the stage and fix it with the slide clips Note The thickness of t...

Страница 15: ...of the field of view At this time stopping the field diaphragm image just short of the maximum field of view may be convenient for centering Adjust and centre the field diaphragm so that it is just ou...

Страница 16: ...icroscope before making any modifications Figure 10a Figure 10b 1 2 3 3 2 Replace the objective Lower the stage all the way with the coarse focus knob Rotate the nosepiece to move the objective to be...

Страница 17: ...eads can be con nected to the photo port of the biological microscope using an adapter a 0 5X adapter is supplied Screw the adapter onto the camera Loosen the setscrew of the photo port and remove the...

Страница 18: ...o the slot Rotate the nosepiece to lock into the slot Condenser is not set properly Set the condenser properly Aperture iris diaphragm is not set accu rately Set the aperture iris diaphragm accurately...

Страница 19: ...rage Permissible environment temperature Permissible relative humidity 10 40 C Below 31 C max humidity is 80 at 40 C linearly decreases to 50 Operation Permissible environment temperature Permissible...

Страница 20: ...nser fixed K hler 140x135mm 75x30mm On the right On the right On the left of movable clip Used for 4X 100X objectives Binocular tube 30 20 Length of mechanical tube Maximum field number Hinge type int...

Страница 21: ...21...

Страница 22: ...ouse 311 Gloucester Road Causeway Bay Hong Kong Tel 852 2837 0888 Fax 852 2882 2792 Motic Spain S L U Spain Pol Ind Les Corts C Les Corts 12 08349 Cabrera de Mar Barcelona Spain Tel 34 93 756 6286 Fax...

Отзывы: