8

2. Setting Up The Instrument 

Avoid placing the instrument in locations exposed to direct sun-
light, dust, vibration, high temperature and high humidity and 
where it is difficult to unplug the power supply cord.

2.1. Operating environment

 Indoor use

 Altitude: Max 2000 meters

 Ambient temperature: 15°C to 35°C

 Maximum relative humidity: 75% for temperature up to 31°C

decreasing linearly to 50% relative humidity at 40°C

 Supply voltage fluctuations: Not to exceed ±10% of the

normal voltage.

 Pollution degree: 2 (in according with IEC60664)

 Installation / Overvoltage category: 2

(in according with IEC60664)

 Air pressure of 75kPa to 106 kPa

 No hoar frost, dew, percolating water, rain

3. Assembling The Microscope

3.1. Verifying Input Voltage

 The automatic voltage selection works with a broad range of 

settings. However, always use a power cord that is rated for the 
voltage used in your area and that has been approved to meet
local safety standards. Using the wrong power cord could cause 
fire or equipment damage.

 In case of using the extension cord, use only a power supply 

cord with a protective earth (PE) wire.

 In order to prevent electric shock, always turn the power switch 

on the power supply off before connecting the power cord.

 Electrical Specifications:

a. Halogen

Input : 90-240V~, 38W, 50-60Hz (Hologen)
Lamp : 6V30W Halogen
Fuse : 250V T2.5A (If the original fuse is blown, please replace 
with specified fuse)

b. LED

Input : 90-240V~, 6W, 50-60Hz (LED)
Lamp : 3W LED
Fuse : 250V T1A (If the original fuse is blown, please replace 
with specified fuse)

3.2. Lamp and Lamp House Cover
(Replacing The Lamp)

The lamp and the lamphouse become very hot during and after 
a period of operation.
Risk of burn – Do not touch the lamp during or immediately 
after period of operation.
Make sure the lamp has cooled sufficiently before attempting 
to replace the lamp.

a. Halogen

 In order to prevent electric shock always turn the power 

switch off and unplug the power cord before installing or re-
placing the lamp.

 Place microscope on its back and pull back the lamp house 

cover plate.

 Firmly insert the lamp into the socket pinholes until it reach-

es the limit. Be careful not to tilt the lamp when mounting.

 When installing the lamp, do not touch the glass surface of 

the lamp with bare fingers. Doing so will cause fingerprints, 
grease, etc., to burn onto the lamp surface, reducing the illu-
mination provided by the lamp. If the surface is contaminated, 

Содержание BA210 series

Страница 1: ...Instructions Manual English Motic Incorporation Ltd UL Listed Product E250223...

Страница 2: ...lies in its ability to accommodate modular accessories in the optical path and produce a flexible design Specimen Objective Tube lens Intermediate image Eyepiece Lens of eye Retina Optical System Usin...

Страница 3: ...eld of View Objective Magnification For example BA210 Eyepiece field of view 20mm Objective magnification 10X Microscope Terminology Diameter of the object field 20 10 2 0mm Diopter adjustment The adj...

Страница 4: ...rements of modern research techniques and testing methods This involves modification to the mechanical structure and optical design of our instruments Therefore all descriptions and illustrations in t...

Страница 5: ...Component 4 1 Coarse And Fine Focusing 4 2 Coarse Focus Torque Adjustment 4 3 Coarse Focus Height Stopper 4 4 Optical Path Change Over Slider 4 5 Interpupillary Distance Adjustment 4 6 Diopter Adjust...

Страница 6: ...iece lock screw Specimen holder Condenser focus knob Coarse focus torque adjustment ring Fine focus knob Coarse focus knob Interpupillary distance scale Field lens Stage X Axis travel knob Stage Y Axi...

Страница 7: ...ce tube Eyepiece tube Clamp Screw Power switch Brightness control knob Power input Fine focus knob Revolving nosepiece Condenser aperture scale Condenser position lock Condenser aperture lever Condens...

Страница 8: ...n order to prevent electric shock always turn the power switch on the power supply off before connecting the power cord Electrical Specifications a Halogen Input 90 240V 38W 50 60Hz Hologen Lamp 6V30W...

Страница 9: ...it board is located behind the back cover plate 3 Disconnect the LED connection cables from the power supply printed circuit board 4 Loosen LED board locating ring take away LED board locating ring 5...

Страница 10: ...s approximately four times greater As long as the lamp voltage is kept constant the halogen lamp maintains the same level of brightness and colour tem perature regardless of whether it is new or neari...

Страница 11: ...Turning the coarse and fine focus knobs further than their limit 4 2 Coarse Focus Torque Adjustment To increase the torque turn the torque adjustment ring located behind the left hand coarse focus kno...

Страница 12: ...and left field of view become one This adjustment will enable the user to observe the specimen with both eyes 4 6 Diopter Adjustment Diopter adjustment compensates for the differences in vision betwe...

Страница 13: ...the front of the objective The immersion oil supplied by Motic is synthetic non fluo rescing and non resining oil with a refractive index of 1 515 Normally cover glass must be used with oil immersion...

Страница 14: ...not used on immersion procedure Air bubbles in immersion oil Specified immersion oil used not used Immersion oil on dry objective Greasy residue on eye lens Incorrect illumination Uneven focus Specim...

Страница 15: ...ctors are not securely connected Lamp near end of service life Lamp not securely plugged into socket Problem Possible Cause No cohesion of binocular image Magnification or field of view of left and ri...

Страница 16: ...orm Store the objectives eyepieces and filters in a container or desiccator with drying agent Proper handling of the microscope will ensure years of trouble free service If repair become necessary ple...

Страница 17: ...Corts Cam del Mig 112 08349 Cabrera de Mar Barcelona Spain Tel 34 93 756 6286 Fax 34 93 756 6287 Motic Deutschland GmbH GERMANY Christian Kremp Strasse 11 D 35578 Wetzlar Germany Tel 49 6441 210 010...

Отзывы: