5. MAINTENANCE & TROUBLESHOOTING

5.1 Maintenance

The stereo microscope is limited to the following maintenance only:

• Turn off the power switch after use, and put on the dustproof cover after the stereo microscope has cooled down.
• Do not operate the stereo microscope in a room with humidity higher than 75%.
• Remove dust or ordinary dirt on optical lens surfaces with a brush, rubber bulb blower and a moistened lens tissue.
• Use only optical lens tissues and optical lens cleaner (see below). Never clean a lens with a dry optical lens tissue. 

Be sure to remove any dust before using lens tissue and cleaner.

• To remove stubborn oily or lipoid dirt (such as fingerprints), moist the lens tissue with a 3 to 7 ethanol-ether mixture  

or a commercially available optical lens cleaning solution and then use it to wipe off the dirt.

• When cleaning an optical lens surface, wipe gently in a circle from the center to the edge.

5.2 Troubleshooting

Common fault

Possible cause and remedy

Illuminator is not lit

Check if the power switch is turned on.

Check the connection between the power cable and the power supply.

Check if the fuse has blown; if yes, replace with the spare fuse or contact  
your local distributor.
Check if the illuminator is burnt out; if yes, replace the illuminator or contact  
your local distributor.
Check if the supply voltage is the same as the working voltage of the instrument 
(100~240VAC 50/60Hz).
If it’s not caused by any of the above causes, contact your local distributor.

Bulb flashes

Loose contact of the illuminator plug socket. Connect the plug socket properly.

Fail to adjust focus  
for specimen

Check if the specimen is so big that the stereo microscope does not have enough 
space to adjust the focus.
Check if the tension adjustment ring is so loose that the body slides down due to 
its deadweight. Tighten the tension ring with the supplied hook wrench until the 
body holds in place but will still move easily when the focus knobs are turned
Check if the eyepiece or objective lenses are dirty. If yes, wipe the lens according 
to the instructions in section 3.2 of Operation of the stereo microscopes.

Poor clarity  
or obscure image

Failure to adjust the focus for the specimen accurately. Focus using the procedure 
outlined in section 3.2.
Dirty objective lens. Wipe the lens according to the instructions in section 5.1  
of the Operation Instructions
Dirty eyepiece lens. Wipe the lens according to the instructions in section 5.1  
of the Operation Instructions

Содержание 1100201000026

Страница 1: ...Motic Incorporation Ltd Instructions Manual English RED30 SERIES...

Страница 2: ......

Страница 3: ...2 2 Part names 08 2 3 Application 09 2 4 Instrument and its technical specifications 09 3 First time use operation 10 3 1 First time use 10 3 2 Operation of the stereo microscopes 12 3 3 Modification...

Страница 4: ......

Страница 5: ...to the EN 61010 1 2001 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use 1 3 Unpacking transportation storage The original shipping container a foam box in a fibe...

Страница 6: ...Motic microscope repair agency to have it repaired Please be sure to turn off the power before opening the instrument to replace LED illuminator or replace the fuse Only use a fuse for the rated curr...

Страница 7: ...7 2 1 General Views 2 INSTRUMENT DESCRIPTION RED39Z...

Страница 8: ...1 Eyepiece 6 Base 2 Binocular tube 7 Handle 3 Objective turret 8 Focus knob 4 Sample clip holder 5 Stage plate 9 Brightness adjustment knob incident illumination 10 Brightness adjustment knob transmit...

Страница 9: ...king distance 47mm 68mm Built in Rechargeable Battery N Y Y Illumination Incident and Transmitted LED illumination Optional Objectives Objectives 1X 3X 4X All of the models in the RED30 Series stereo...

Страница 10: ...of the eyepieces and other optical components do not touch the lens surface Make sure the lock screws of the eyepieces are adjusted correctly The eyepieces should turn easily but should not be able to...

Страница 11: ...ent LED brightness at the right hand of the base Figure 3a Control the transmitted LED brightness at the left hand of the base Figure 3b After use turn the brightness control to the minimum position a...

Страница 12: ...models Observe with the right eyepiece only and rotate the focus knob until the image is in focus Observe with the left eyepiece and adjust the diopter ring on the left body tube until a focused imag...

Страница 13: ...aining ring from the LED array and then unplug remove and finally replace the LED array When replacing the Incident illumination for RED39Z first remove the turret cover glass and the ring and then Sl...

Страница 14: ...ents the eyepiece from falling out The diopter ring is located on the left body tube Please refer to the Section 3 2 2 of the Operation Instructions for the correct method of adjustment 4 2 Stand The...

Страница 15: ...place with the spare fuse or contact your local distributor Check if the illuminator is burnt out if yes replace the illuminator or contact your local distributor Check if the supply voltage is the sa...

Страница 16: ...rature Permissible relative humidity 10 40 C Below 31 C max humidity is 80 at 40 C linearly decreases to 50 Operation Permissible environment temperature Permissible relative humidity 5 40 C Below 31...

Страница 17: ...cus adjustment mechanism Coarse focus adjustment Stroke 19 8mm rotation 47mm Objective RED30S RED39Z Paired 2X and 4X 1X 3X Eyepiece Field number 20mm Assembly diameter 30mm W10X 20 In the registered...

Страница 18: ......

Страница 19: ......

Страница 20: ...ouse 311 Gloucester Road Causeway Bay Hong Kong Tel 852 2837 0888 Fax 852 2882 2792 Motic Spain S L U Spain Pol Ind Les Corts C Les Corts 12 08349 Cabrera de Mar Barcelona Spain Tel 34 93 756 6286 Fax...

Отзывы: