background image

12

4.8. Use Of Aperture Diaphragm

• The condenser aperture diaphragm is provided for adjusting 
the numerical aperture (N.A.) of the illuminating system of the 
microscope, it decides the resolution of the image, contrast, 
depth of focus and brightness.
• Stopping down will lower the resolution and brightness but 
increase the contrast and depth of focus. 
• An image with appropriate contrast can be obtained with an 
aperture setting that is 2/3 of the objective N.A.
• To adjust the aperture diaphragm:
- adjust the condenser aperture diaphragm ring referring to the 
condenser aperture scale, or
- by observing the diaphragm image visible on the exit pupil 
inside the eyepiece tube, or
- by using a centering telescope after removing one of the eye-
pieces and focusing on the aperture diaphragm.

4.9. Use of field diaphragm

• The field diaphragm determines the illuminated area on the 
specimen. Rotating the field diaphragm ring changes the size of 
the field diaphragm. For normal observation, the diaphragm is 
set slightly larger than the field of view. If a larger than required 
area is illuminated, extraneous light will enter the field of view. 
This will create a flare in the image and lower the contrast.
• The thickness of the glass slide must be 1.7mm or less, otherwise 
the field diaphragm may not be focused on the specimen plane. 
• The diaphragm does not have any effect when the condenser 
top lens is swung out of the optical path in the Swing-out type 
condenser. Fully open the field diaphragm, as the N.A. of the
illuminating system will be reduced if the diaphragm is exces-
sively stopped down.

4.10. Brightness and contrast adjustment

• Neutral density filters are used for brightness adjustment in 
routine microscopy and photomicrography.
• Green interference (546nm) for phase contrast and contrast 
adjustment with black and white film.
• HE (didymium filter) for colour photomicrography of the 
Haematoxylin & Eosin (HE) or Fuchsin stained specimen with 
tungsten type film.

4.5. Interpupillary Distance Adjustment

• Before adjusting the interpupillary distance, bring a specimen 
into focus using the 10x objective.
• Adjust the interpupillary distance so that both the right and 
left field of view become one.
• This adjustment will enable the user to observe the specimen 
with both eyes

4.6. Diopter Adjustment

• Diopter adjustment compensates for the differences in vision 
between the left and right eyes. In addition to making observa-
tion through both eyes easier, this adjustment also reduces the 
extent to which focusing is lost when the objective magnifica-
tion is changed. In particular, this occurs when a low magnifi-
cation objective is used.
• Before adjusting the diopter, bring a specimen into focus us-
ing the 10x objective.
• Turn the diopter compensation ring on each eyepiece until 
the adjustment ring is adjusted to the “0” position.  Position 
the 40x objective into the optical path and bring the specimen 
image into focus by turning the coarse and fine focus knobs.
• Position either the 4x or 10x objective into the optical path. 
Without adjusting the fine and coarse focus knobs, turn the 
diopter rings on the eyepieces so that the specimen images in 
the left and right eyepieces are focused individually.
• Repeat the above step twice.

4.7. Centering the condenser

• Fully open the field of view diaphragm and condenser aper-
ture diaphragm.
• Set the specimen on the stage with the cover glass facing up.
• Bring the specimen image into focus, using the 10X objective.
• Close the field of view diaphragm to its minimum setting by 
means of the field diaphragm ring.
• Turn the condenser focus knob to bring the field diaphragm 
image into focus on the specimen plane.
• Adjust the condenser centering screws so that the image of 
the field diaphragm appears at the centre of the field of view. 
At this time, stopping the field diaphragm image, just short of 
the maximum field of view, may be convenient for centering.
• Adjust and centre the field diaphragm so that it is just out-
side the field of view for each magnification change. 

Содержание 1100100401483

Страница 1: ...Motic Incorporation Ltd UL Listed Product E250223 Instruction Manual English...

Страница 2: ...ements of modern research techniques and testing methods This involves modification to the mechanical structure and optical design of our instruments Therefore all descriptions and illustrations in th...

Страница 3: ...he light reaching the objective originates from the periphery of the specimen and enters the optical system at oblique angles moving forward diagonally but still in parallel bundles toward the tube le...

Страница 4: ......

Страница 5: ...4 1 Coarse and fine focusing 4 2 Coarse focus torque adjustment 4 3 Coarse focus Knob lock 4 4 Optical path change over slider 4 5 Interpupillary distance adjustment 4 6 Diopter adjustment 4 7 Center...

Страница 6: ...is travel knob Stage Y Axis travel knob Objective Binocular eyepiece tube Eyepiece Diopter adjustment ring Eyepiece lock screw Stage Condenser focus knob Coarse focus torque adjustment ring Fine focus...

Страница 7: ...scale Condenser centering screws Specimen Holder Condenser clamp screw Photo port Optical Path Changeover lever Trinocular eyepiece tube Eyepiece tube Clamp Screw Power switch Brightness control knob...

Страница 8: ...er to prevent electric shock always turn the power switch on the power supply off before connecting the power cord Electrical Specifications a Halogen Input 90 240V 80VA 50 60Hz Lamp 6V 30W Halogen Fu...

Страница 9: ...t board is located behind the back cover plate 3 Disconnect the LED connection cables from the power sup ply printed circuit board 4 Loosen LED board locating ring take away LED board locat ing ring 5...

Страница 10: ...dless of whether it is new or nearing the end of its life b LED This is the first 3W microscopic illumination system with a global patent for long life adjustable high intensity heat gen eration and e...

Страница 11: ...while holding the other Turning the coarse and fine focus knobs further than their limit 4 2 Coarse Focus Torque Adjustment To increase the torque turn the torque adjustment ring located behind the le...

Страница 12: ...E or Fuchsin stained specimen with tungsten type film 4 5 Interpupillary Distance Adjustment Before adjusting the interpupillary distance bring a specimen into focus using the 10x objective Adjust the...

Страница 13: ...the front of the objective The immersion oil supplied by Motic is synthetic non fluo rescing and non resining oil with a refractive index of 1 515 Normally cover glass must be used with oil immersion...

Страница 14: ...lution Condenser is set too low Aperture diaphragm closed too far No cover glass Too thick or thin cover glass Immersion oil not used on immersion procedure Air bubbles in immersion oil Specified imme...

Страница 15: ...amp near end of service life Lamp not securely plugged into socket Problem Possible Cause Insufficient parfocality of objectives Eyepiece diopter not adjusted No cohesion of binocular image Magnificat...

Страница 16: ...dirt dampen a piece of gauze with diluted neu tral detergent and wipe lightly C Disinfecting the Microscope Follow the standard procedures for your laboratory D When not in use When not in use cover t...

Страница 17: ...l Mig 112 08349 Cabrera de Mar Barcelona Spain Tel 34 93 756 6286 Fax 34 93 756 6287 Motic Deutschland GmbH GERMANY Christian Kremp Strasse 11 D 35578 Wetzlar Germany Tel 49 6441 210 010 Fax 49 6441 2...

Отзывы: