![Morgan MEDesign BASIC-ONE CSA-UL Скачать руководство пользователя страница 9](http://html.mh-extra.com/html/morgan/medesign-basic-one-csa-ul/medesign-basic-one-csa-ul_maintenance-and-service-manual_1833851009.webp)
DIAGNOSTICS
NOISY MOTOR
GREEN STATUS LIGHT ON?
ALARM STATUS LIGHTS ON?
Consider to
REPLACE MOTOR
CHECK AC MAINS for low voltage
MONITOR CIRCUITS are O.K.
FILTER is O.K.
FAIRCHILD MONITOR is O.K.
LINE VOLTAGE is O.K.
+12 VOLT SUPPLY is O.K.
Is the MOTOR OVERHEATED?
NO YES
WEIGHT TOO GREAT?
NO Check limits.
All MOTOR mechanical
connections secure and in place?
YES
Morgan MEDesign
, Inc. - Maintenance ~ Service Manual BASIC-ONE -
2005 Page 8
POSSIBLE OVERHEATED MOTOR
The motor drives have a duty cycle and
when this is exceeded will self-protect
with a thermal electrical breaker. WAIT
10 MINUTES and TRY OPERATION
AGAIN
WAIT 10 MINUTES MORE,
if NO OPERATION consider
to REPLACE the AFFECTED
MOTOR