Voltage dips,
short interrup-
tions and voltage
variations on
power supply
input lines
IEC61000-4-11
<5 % UT (>95
% dip in UT)
for 0,5 cycle
40 % UT (60
% dip in UT)
for 5 cycles
70 % UT (30
% dip in UT)
for 25 cycles
<5 % UT (>95
% dip in UT)
for 5 sec
<5 % UT (>95
% dip in UT)
for 0,5 cycle
40 % UT (60
% dip in UT)
for 5 cycles
70 % UT (30
% dip in UT)
for 25 cycles
<5 % UT (>95
% dip in UT)
for 5 sec
Mains power quality should be
that of atypical commercial or
hospital environment. If the
user of this device requires
continued operation during
power mains interruptions, it is
recommended that the device
be powered from an uninterrup-
tible power supply or a battery.
Power frequency
(50/60Hz)
magnetic field
IEC61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of atypical
location in a typical commercial
or hospital environment.
NOTE
UT is the a.c. mains voltage prior to the application of the test level
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity:
Thie Supra devices are intended for use in the electromagnetic environment specified below.
The user of this device should make sure it is used in such an environment.
Immunity Test
IEC60601
test level
Compliance
Electromagnetic
Environment-Guidance
Conducted RF
IEC 61000-4-6
3Vrms150 kHz to
80 MHz outside
ISM bandsa
3 Vrms
Portable and mobile RF com-
munications equipment should
be used no closer to any part
of this device, including cables,
than there commended sepa-
ration distance calculated from
the equation applicable to the
frequency of the transmitter.
Recommended
separation distance
d = 1.2
Î
P 150 kHz a 80 MHz
d = 1.2
Î
P 150 kHz a 80 MHz
d = 2.3
Î
P 80 MHz a 2.5G
MHzWhere P is the maxi-
mum output power rating of
the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).b
Field strengths from fixed
RF transmitters, as determi-
ned by an electromagnetic
site survey c, should be less
than the compliance level in
each frequency ranged.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Radiated RF
IEC 61000-4-3
3 V/m 80 MHz
to 2.5 GHz
3 V/m