ENGLISH
26
-Static Mode
If the pump unit is equipped with STATIC Low Air Loss only function, press the orange button
to suspend the alternating mode of mattress. By pressing this button again, the mattress will
be back to alternating therapy mode.
-Fuse Replacement
1.
Disconnect the plug form mains power when a blown fuse is suspected
2. Remove the cover of the fuse holder by means of a small screwdriver
3. Insert a new fuse of the correct rating in, and replace the cover of the fuse holder back
The fuse should be rated as T1AL/250V type
STORAGE
• Fold the mattress accurately to prevent creases
• Roll from the head end towards the foot end with the valve open
MAINTENANCE
The LEVITAS devices by MORETTI are checked carefully and supplied with EC mark, once
launched on the market. For patient and physician’s safety we recommend to let check the
device by the manufacturer or by laboratory authorized every year. In case of repair, use
only original spare parts and accessories.
CLEANING AND DISINFECTION
Before using the device with people follow this procedure:
• Clean the mattress with mild soap
• Do not immerse or soak pump unit
• After cleaning, dry the mattress without direct sunlight exposure.
• Check the connection tubes are not obstructed by dust or else. In this case use a soft swab
to clean the tubes
NOTE
Never use acids, alkalis or solvents such as acetone
WARNING!
• Unplug the pump before starting the cleaningprocedure