11.1 Warning for the battery
1. Keep the battery out of the reach of children
2. The battery is composed by lithium batteries, improper use may cause explosion
3. Dispose the battery following the local procedures
12. HOW TO USE
12.1 Front Panel
Therapy Modes
To switch the therapy mode:
A. Alternating Mode
Continuously and sequentially inflate and deflate air cells to remove constant pressure and
facilitate improved cellular perfusion.
Static Mode
Provides a stable all cell inflated surface at lower pressure when compared to the respective comfort
level in alternative mode. The system then redistributes patient weight over a greater surface.
Cycle Time
To adjust the pressure alternating cycle time to 10, 15 or 20 minutes.
Comfort Setting
To adjust the comfort level according to user’s preference or doctor’s advice. There are 5 different
comfort levels: 1 is the softest & 5 is the firmest. The default setting is at comfort level 1.
NOTE.
A hand check is needed to determine if patient is bottoming out. When a patient’s condition
has significantly changed reassess appropriateness of product and comfort setting level.
Alarm Mute
Press alarm mute button to temporary suspend the alarms. Should the problem continue, the
alarms will resound within 10 minutes to notify the caregiver until resolved.
5. Low Battery Indicator
A low battery alarm indicates approximately 1 hours of therapy remain; charge batteries
IMMEDIATELY to prevent disruption of therapy. When battery is charged, audible and visual
alarms will turn off.