CHAPTER 17 Trouble shooting   

Problems 

Possible reason 

Solution 

SpO

2

% or pulse rate 

can  not  be  shown 
normally 

1. Finger is not plugged correctly 
2. Patient’s SpO

value is too low 

to be measured 

1. Retry by plugging the finger 
 
2.  Measure  more  times,  If  you  can 
make sure about no problem existing 
in  the  product.  Please  go  to  a 
hospital timely for exact diagnosis 

SpO

2

% or pulse rate 

is shown unstably 

1.  Finger  may  not  be  plugged 
deep enough 
2. Excessive patient movement 

1. Retry by plugging the finger 
2. Be calmness. 

The Monitor can not 
be powered on 

1.  No  battery  or  low  power  of 
battery 

2.  Battery  might  be  installed 
incorrectly 

3.  The  Monitor  might  be 
damaged 

1. Please replace battery 

2. Please reinstall the battery 

3. Please contact with local 
customer service center 

Displaying 

screen 

can 

not 

be 

refreshed 

1. SpO

  module damaged 

2. Hang     

1.  Contact  with  local  customer 
service center       

2.  Re-power  on  (cut  off  the  AC 
power and fetch out DC battery ) 

 
 
 
Blood pressure can 
not be detected and 
displayed.   

1. 

The 

body 

part 

being 

measured is not proper 

2.  The  patient  is  measured  with 
coat 

3. The patient moves and speak 
during measurement 

4. The device is operated in EMI 
circumstance 

 

1. Make sure the body part wrapped 
by    cuff is correct. 

2.  Patient  should  make  upper  arm 
bared, and closely touched with cuff. 

3.  Patient  should  keep  quiet  and 
motionless. 

4.  Move  to  another  place  far  away 
form  Interference  source   such  as 
bedroom.   

 

Data storage failure    1.  The ID number is zero 

2.  The memory is damaged 

1.  Set  the  ID  number  in  range  of  “ 

1~99”   

2.  Please  contact  with  the  local 

customer service center 

  Printing is in 
trouble. 

1. Low power   

2. Wrong operation   

1.  Please  change  the  battery  or 
adopt the AC power supply 

2.  Please  refer  to  the  instruction 
manual for printing. 

 

Others 

 

Contact      the  technician  in  Service 
Department of our company. 

Содержание Dimed

Страница 1: ...MA LTD850 MOD MD2000B 02 A_06 2015 TABLE PULSEOXIMETER INSTRUCTION MANUAL Coperta LTD850 01 A indd 1 25 08 14 10 07...

Страница 2: ...s work unless expressly authorized by our company All information contained in this publication is believed to be correct We shall not be liable for errors contained herein nor for incidental or conse...

Страница 3: ...rted 15 4 1 Unpacking and Inspection 15 4 2 Connect the Power Cables 16 4 3 Power on the Monitor 16 4 4 Connect with Patient Sensors 16 4 5 Check the recorder 16 CHAPTER 5 Measuring screen and Main me...

Страница 4: ...iority 30 10 2 Alarm set 31 CHAPTER 11 Real Print Set Optional 32 CHAPTER 12 Network Set Optional 32 CHAPTER 13 System Set 33 CHAPTER 14 System Configuration 34 CHAPTER 15 NIBP Operation 34 CHAPTER 16...

Страница 5: ...Therefore assure proper venous outflow from monitored site Sensor should be not below heart level e g sensor on hand of a patient in a bed with arm dangling to the floor Avoid placing the sensor on an...

Страница 6: ...system is in correct working state and operating condition As with all medical equipment carefully route patient cabling to reduce the possibility of patient entanglement or strangulation Do not use...

Страница 7: ...ully read the relevant chapter in this instruction FEDERAL LAW U S A restricts this device to sale by or on the order of a physician 1 4 Explanation of Symbols Type BF applied part Power on off button...

Страница 8: ...dult and pediatric patients in hospitals medical facilities and sub acute environments The vital signs monitor is intended for spot checking and or continuous monitoring of patients CHAPTER 2 Overview...

Страница 9: ...Down button Press this button you can select different item and decrease the number 11 CHG indication lamp Battery charge indicator If the battery is charging the lamp will flash When the battery is c...

Страница 10: ...cket is connected with the center monitor system of our company The function is reserved 3 I O for programme upgrade The function is reserved 4 Speaker 5 Power socket AC Power supply socket 6 Ground t...

Страница 11: ...ughly once a month If the battery low indication icon displays during measuring please connect the monitor to AC power socket timely avoiding the current measuring is affected Warning Do not use the b...

Страница 12: ...requirement Only standard 50 0 1 mm thermosensitive record paper can be used otherwise the recorder may not function the recording quality may be poor and the thermosensitive print head may be damage...

Страница 13: ...lowing conditions Operating Temperature 5 to 40 Relative Humidity 80 non condensing Standard Atmosphere Pressure 86kPa 106kPa The transport and storage environment should comply with the following con...

Страница 14: ...g Range Parameters Upper alarm range Low alarm range SpO2 71 100 70 99 PR bpm 31 254 30 253 SYS bpm 61 255 60 220 DIA bpm 31 220 30 180 MAP bpm 31 220 30 200 Outline Dimension 296mmX166mmX96mm Weight...

Страница 15: ...ing operation the instrument is free of condensation Condensation can form when equipment is moved from one building to another thus being exposed to moisture and differences in temperature 3 2 Ground...

Страница 16: ...After the probe is connected to its input cables turn the monitor on by momentary pressure on the front panel POWER button for 3 seconds As audible beep feedback after pressing the monitor is initiate...

Страница 17: ...008 08 25 18 00 Silence audio alarm indicator the current audio alarm status is on Battery power indicator 2 The current system status in fig 11 USB function indicator status On Net function indicator...

Страница 18: ...larm limits PR alarm limits indication and measurement information The current SpO2 alarm limits Upper alarm 100 Lower alarm 95 The current PR alarm limits Upper alarm 120 Lower alarm 60 The current m...

Страница 19: ...ngs After entering into menu 2 press the button to select Date And Time item and press the button to confirm your selection you will see the following picture Fig 14 Press the button and button to sel...

Страница 20: ...99 CHAPTER 7 Take a Measurement 7 1 SpO2 and PR measurement After the time and the ID number settings connect the sensor and take a measurement according to the following pictures Firstly select the s...

Страница 21: ...s Other SpO2 sensors may cause improper performance 4 Do not use a SpO2 sensor with exposed optical components 5 Excessive patient movement may cause inaccurate measurements 6 Tissue damage can be cau...

Страница 22: ...f measurement available manual Auto and STAT Each mode displays the diastolic systolic and mean blood pressure WARNING You must not perform a NIBP measurement on patients with sickle cell disease or u...

Страница 23: ...r company If you are not familiar with this monitor and NIBP parameters then follow this chapter sections 1 Plug in the extended tube and switch on the system Ensure the luer lock to plug into the cor...

Страница 24: ...Body blood pressure is of volatility character therefore you should take several measurements in several days to judge whether or not the blood pressure is duratively increased Detailed measurement re...

Страница 25: ...onged cycling of measurement serious tremble will lead to the failure of measure b Placing the cuff too loosely or tightly on the patient c Leaky cuff or hose d Insure the NIBP and pulse rate within t...

Страница 26: ...us screen Fig 18 1 Fig 18 2 Select ID with 1 99 IDs selectable for data reviewing Start Print By the item you can print the records of the selected ID Stop Print By the item you can stop the print act...

Страница 27: ...ID 8 3 PR Trend Review You can review the PR trend of the current ID or a different ID PR trend by setting manually In menu 1 Press the button to select the PR Trend review item and then press the bu...

Страница 28: ...ation is similar to the 8 2 SpO2 Trend Review refer to section 8 2 Fig 22 1 Fig 22 2 8 6 Erase Data Select the Erase Data item in menu 2 and then press the button to enter into the Erase Data screen r...

Страница 29: ...rogress will stop Manual mode under the mode press the NIBP button you can start a single measuring STAT mode the monitor will measure the blood pressure continuously in 5 minutes After completion of...

Страница 30: ...alarm is activated by more than one physiological alarm each parameter will be displayed flashing with striking color meanwhile the alarm information is displayed in information column circularly AUD...

Страница 31: ...he default SpO2 low alarm is 95 PR Upper alarm range 31 254 The default PR upper alarm is 120 Low alarm range 30 253 The default PR low alarm is 50 SYS Upper alarm range 61 255 The default PR low alar...

Страница 32: ...a with the range of 4 40s Numbers for selection contain 4 8 12 16 20 24 28 32 36 40 Auto Print Start Print By the button the printer begins to print the records after about 3 seconds with the change o...

Страница 33: ...ion is reserved currently Alarm Volume Press the or button to pick the Alarm Volume sub item and press the button to confirm Press the or the button to increase or decrease the value of alarm volume A...

Страница 34: ...d configuration Restore configuration you have selected Save Current configuration Save the current settings made by the current user Factory Default initial settings made by manufacturer User Saved t...

Страница 35: ...ying screen can not be refreshed 1 SpO2 module damaged 2 Hang 1 Contact with local customer service center 2 Re power on cut off the AC power and fetch out DC battery Blood pressure can not be detecte...

Страница 36: ...p bottom and front surfaces lightly The following admissive liquor can be used Ammonia diluted Glutaraldehyde Sodium hypochlorite bleacher diluted Mildness suds diluted Please obey the following rules...

Страница 37: ...al repair team trained by our company provide this service Fittings replacement if necessary we will replace the damaged fittings according to contract The damaged fittings should be returned to us ex...

Страница 38: ...ion with result form improper repair by anyone other than our company authorized service personnel The malfunction with result from improper use 18 2 5 Special warranty period required by Customer We...

Страница 39: ...e cuff for adult RNC0001A 013B 1 piece 9 3 meter metal head black cuff extended tube 1 piece The optional accessories No Accessories Quantity 1 SpO2 probe for adult Model M 50G19CSO 1 piece 2 SpO2 pro...

Страница 40: ...Via Bruxelles 3 Meleto 52022 Cavriglia Arezzo Tel 39 055 96 21 11 Fax 39 055 96 21 200 www morettispa com info morettispa com...

Отзывы: