Mopec 5000 Скачать руководство пользователя страница 15

.01&$

 5000 SCHEMATIC WIRING

Medezine Limited
Churchill Way  Chapeltown 
Sheffield  England
S35 2PY 

T
F +44 (0)114 257 6222
E   [email protected]
W www.medezinesaws.com

+44 (0)114 257 6002

2002/289

ST

AR

T/STOP

MOTOR

PRINTED

CIRCUIT

SA

W

MOTOR

SA

W UNIT

Connections

shown at 

opposite end of

power supply

enclosur

e for

clarity

A6

A5

A4

A3

I/O

I/O

M +

M -

SA

W

PL5

1

2

3

4

5

EAR

TH

1

2

3

4

5

6

LIVE!

LIVE!

L

 Fused

N Fused

L

N

Mains IN

Pump Power OUT

PL1

PL2

2

1

2

1

F2

F1

V

aristor

250V

220/230

S1

110/1

15

-V

VR1

RL1

F3

PL4

T

ransformer secondaries

EAR

TH

1

2

3

4

5A

 QB

D1

KBU4D

Fusible Resistor

D4

S1PB-E3/84A

F4

22R

IN  OUT

GND

Pump Control

Saw Control

5V Regulator

U1

Mains Supply (Green)

+C1

4700uF

63V

40V

V

acuum (Green)

+C3

22uF

35V

0V

1

3

2

5V

0V

2

2

C5

100uF

25V

C4

100uF

25V

0V

Saw 40V

0V

C6

100uF

25V

1

1

OVERLOAD (Y

ellow)

R1

1

10k

C2

100uF

100V

R12

10k

R1

10k

R4

3k9

R5

100R

R6

0R1

R10

1k

R9

1k

0.5W

5V

2

1

SA

W

PL3

T

ransformer Primaries

Ther

mal F

use Out

P1*

P1

Ther

mal F

use In

P2*

P2

R8

10k

On/Of

f

Q3

BCX56-16

RL2

D8

S1PB-E3/84A

R4

3k9

0.5W

D5

m m

D2

m m

2

10k

R7

S1PB-E3/84A

D7

m m

D3

D9

20V

Q2

BCX56-16

D6

m m

1

2

3

4

8

PIC1 2F675

Vcc

GP5

GP4

GP3

Vss

AN0

GP1

GP2

IC1

6

5

7

!

!

!

!

!

!

Saf

et

y C

ritic

al P

ar

ts

Содержание 5000

Страница 1: ...s are made available to the operator and qualified maintenance staff at all times The Mopec 5000 System has been examined and found to be in compliance with the standard for Medical Electrical Equipment IEC EN60601 1 www mopec com 800 362 8491 info mopec com 05 OWFS UJHIUFOJOH PG the Captive Blade Clamp XJMM DBVTF the CMBEF UP OPU TFBU QSPQFSMZ BOE the TBX not to PQFSBUF QSPQFSMZ ...

Страница 2: ... 8BSOJOHT FOFSBM FTDSJQUJPO 4FRVFODF PG 0QFSBUJPO PQFD 4BX 6OJU PQFD 1PXFS 4VQQMZ 1PSUBCMF PQFD 7BDVVN 6OJU JUUJOH DDFTTPSJFT BOE FUBDIBCMF 1BSUT 1SFWFOUJWF BJOUFOBODF OE OTQFDUJPO PQFD VTF 3FQMBDFNFOU PQFD 1PXFS 4VQQMZ 5SPVCMFTIPPUJOH VUUJOH JOUT JTJOGFDUJPO BOE 4UFSJMJ BUJPO DDFTTPSJFT BOE FUBDIBCMF 1BSUT 4DIFNBUJD 8JSJOH JBHSBN JMUFS 4FSWJDF 3FDPSE POEJUJPOT PG 8BSSBOUZ MFDUSJDBM 5FTU FSUJGJDBU...

Страница 3: ... BDDPSEBODF XJUI UIF GPMMPXJOH DPEF 306 653 7 Medezine Limited Churchill Way Chapeltown Sheffield England S35 2PY T F 44 0 114 257 6222 E sales medezinesaws com W www medezinesaws com 44 0 114 257 6002 2002 289 W W W SFFO 8IJUF MBDL SFFO FMMPX MVF SPXO Type B Equipment Protection against electrical shock Saw Unit Watertight Equipment IP67 Power Supply Splash Proof Equipment IP65 EQUIPMENT CLASSIFI...

Страница 4: ...JQNFOU CFGPSF VTF 6TF UIF BQQSPQSJBUF TBGFUZ FRVJQNFOU P OPU JNNFSTF UIF 1PXFS 4VQQMZ BOE 7BDVVN VOJU JO JRVJE 6TF HPPE RVBMJUZ EJTJOGFDUBOUT BT BQQSPWFE CZ UIF PTQJUBM OGFDUJPO POUSPM P OPU VTF DPSSPTJWF DMFBOJOH NBUFSJBMT PS ZQPDIMPSJUF BT UIFTF XJMM EBNBHF UIF VOJUT P OPU MFBWF UIF BOE 1JFDF TPBLJOH JO DMFBOFS GPS QSPMPOHFE QFSJPET PG UJNF BT XJMM DBVTF BO JOHSFTT PG GMVJE 4 5 8 3 4 PUF OWFS UJ...

Страница 5: ...usage of the saw unit This unit is watertight User preference may require the equipment to be supplied in its basic form which comprises the saw unit and power supply only A filtration unit may be easily added by the customer at a later stage The complete MPQFD 5000 system is capable of handling health endangering dusts making it suitable for use in hospitals and industrial environments where the ...

Страница 6: ... I O Momentary Button Watertight 3TRAIN 2ELIEF Saw Unit Suppy Cable M01 5000 SAW UNIT Medezine Limited Churchill Way Chapeltown Sheffield England S35 2PY T F 44 0 114 257 6222 E sales medezinesaws com W www medezinesaws com 44 0 114 257 6002 2002 289 ...

Страница 7: ...Unit Output energized when lit 4UBUVT OEJDBUPS 1PXFS 4VQQMZ FOFSHJ FE XIFO MJU Status Indicator Saw Unit overload detected when flashing Complete system will shut down Then Auto reset to System off status in approx 20 seconds AINS NPUT EAD 7ATERPROOF ONNECTOR 3CREW OCKING ON NTERCHANGEABLE AND 0IECE EAD 7ATERPROOF ONNECTOR 3CREW OCKING ON NTERCHANGEABLE 01 VUPQTZ 4BX 1PXFS 4VQQMZ 1PSUBCMF 0OWER 3U...

Страница 8: ...le Clamp Vacuum Cowelling Filtration Unit Filtration Unit Chassis Supply Lead Filtration Unit Handle Cable Clip PoweR 3UPPLY NCLOSURE Status Indicator lights Medezine Limited Churchill Way Chapeltown Sheffield England S35 2PY T F 44 0 114 257 6222 E sales medezinesaws com W www medezinesaws com 44 0 114 257 6002 2002 289 01 VUPQTZ 4BX 7BDVVN 6OJU 0QUJPOBM 1PXFS 4VQQMZ OQVU FBE ...

Страница 9: ... INSPECTION THE EQUIPMENT SHOULD BE DISCONNECTED FROM THE ELECTRICAL SUPPLY WHILE THE FOLLOWING TASKS ARE BEING PERFORMED THE SAW UNIT Daily Clean thoroughly with recommended disinfectant using soft nylon brush not wire or hard bristle type Rinse down and dry thoroughly Weekly Lubricate retaining screw blade clamp and square drive with light engineering or silicone based grease Graphite based grea...

Страница 10: ...er is seated properly Medezine Limited Churchill Way Chapeltown Sheffield England S35 2PY T F 44 0 114 257 6222 E sales medezinesaws com W www medezinesaws com 44 0 114 257 6002 2002 289 Status Indicator Filtration Unit Output energized when lit Captive Retaining Screws Enclosure Cover Status Indicator Saw Unit Output energized when lit 4UBUVT OEJDBUPS 1PXFS 4VQQMZ FOFSHJ FE XIFO MJU Status Indica...

Страница 11: ...FOU SBUFE WPMUBHF JT DPNQUBCJMF XJUI UIF FMFDUSJDBM TVQQMZ Disconnect filtration unit connector from the power supply If fuses do not blow when saw unit is switched on then the filtration unit is defective Filtration unit has been used to extract liquids System shuts down repeatedly Overload indicator is on Distorted blade Blunt blade Incorrect fitted blade Discard CMBEF immediately Index blade so...

Страница 12: ...r Solution Symptom Faulty connection Excessive noise from saw unit Distorted blade Badly fitting or loose blade RFQMBDF blade Check blade is fitted in accordance with the manufacturers instructions Contact manufacturer Bearing service required Medezine Limited Churchill Way Chapeltown Sheffield England S35 2PY T F 44 0 114 257 6222 E sales medezinesaws com W www medezinesaws com 44 0 114 257 6002 ...

Страница 13: ...on or sterilization method Autoclaving of this equipment is not recommended Only use cleaning agents that are safe for use on metals We do not recommend the use of HyQPDIMPSJUF or corrosive cleaners as possible corrosion could result We advise gentle scrubbing with a soft non metallic brush rinsing with clean water and drying where necessary Do not direct high pressure water jets at any part of th...

Страница 14: ...02 2002 289 7 7 4BX 1PXFS 4VQQMZ 7 7 1PXFS 4VQQMZ BOE 1JFDF POF VTU PMMFDUPS 7 4BX BOE 1PXFS 4VQQMZ 7 POF Bone VTU PMMFDUPS 240V Dust Collector Filter Service Kit Segmented MBEF Circular MBEF 2 5 Segmented MBEF 3 Circular Blade Nut Driver Depth Gauge 64mm Depth Gauge 76mm Vacuum Hoods 10Pack Vacuum Pipe with QR Fittings 1BQFS GJMUFST QBDL PG MPUI GJMUFS JDSPO GJMUFS 5 Captive Blade Clamp Kit Inc A...

Страница 15: ...V VR1 RL1 F3 PL4 Transformer secondaries EARTH 1 2 3 4 5A QB D1 KBU4D Fusible Resistor D4 S1PB E3 84A F4 22R IN OUT GND Pump Control Saw Control 5V Regulator U1 Mains Supply Green C1 4700uF 63V 40V Vacuum Green C3 22uF 35V 0V 1 3 2 5V 0V 2 2 C5 100uF 25V C4 100uF 25V 0V Saw 40V 0V C6 100uF 25V 1 1 OVERLOAD Yellow R11 10k C2 100uF 100V R12 10k R1 10k R4 3k9 R5 100R R6 0R1 R10 1k R9 1k 0 5W 5V 2 1 S...

Страница 16: ...replaced 2 Date of Service Signed Vaccum Pipe replaced Cloth Filter replaced Micron Filter replaced 3 Date of Service Signed Vaccum Pipe replaced Cloth Filter replaced Micron Filter replaced 4 Date of Service Signed Vaccum Pipe replaced Cloth Filter replaced Micron Filter replaced 5 Date of Service Signed Vaccum Pipe replaced Cloth Filter replaced Micron Filter replaced 6 Date of Service Signed Va...

Страница 17: ...licy of MPQFD is one of continuous product development and ensuring our customers the best possible service To achieve this specifications and details may be subject to change without prior notice Equipment owners outside the United 4UBUFT should consult the distributor agent from whom they purchased the equipment regarding spare parts and repair facilities The distributor agent will advise on the...

Страница 18: ...ure to IEC529 DIN40050 Protected to IP65 Protection Electronic Current Limiting Device Double Pole Fused Constructed to IEC60601 1 Weight 1 7kg Input voltage 230Vac or 110 Vac Cable 4 0m fitted Hand Piece Motor DC Permanent magnet Speed no load 13200 cycles per min Current no load 40Vdc 0 75amp Typical load 1 5 amps Construction Anodised Aluminium body with 5 bearing sealed mechanism Protected to ...

Отзывы: