T2X
Installing and Connecting the T2X
Now that the T2X has been programmed to your satisfaction, it is ready for installation. The T2X is available
in an HPP housing style, but may be enclosed in any one of several types of certified or uncertified connection
heads. Consult your local Moore Industries’ Interface Solutions Expert for information on the type of enclosure
that best suits your application.
2-WIRE RTD
or DECADE
RESISTANCE
BOX
+PS
2
3
4
1
–PS
–
–
+
+
–
+
POWER
SOURCE
8-42Vdc
CURRENT-
DRIVEN
INSTRUMENT
3-WIRE RTD
OR DECADE
RESISTANCE BOX
4-WIRE RTD
OR DECADE
RESISTANCE BOX
POTENTIOMETER
INPUT
2
3
4
1
2
3
4
1
2
3
4
1
2
3
4
1
2-WIRE RTD
OR DECADE
RESISTANCE BOX
Figure 3. Connecting your T2X to the loop
.
Connecting the T2X in an Application
Moore Industries suggests installing the T2X by first mounting the unit in its intended application, then making
the electrical connections to input, output, and power. Before any installation, make sure that the unit has been
bench checked to ensure that it is configured and calibrated properly for its intended application.