3.7 V 1000 mAH Built-in rechargeable
li-polymer battery, 5V 1A AC Adapter
Digital LCD display
52.00mm diameter circular
Approx. 265 g
Approx. 214 mm × 74.5 mm × 42 mm
A01
Power supply
Display mode
Measurement mode
Oscillographic testing mode
Measurement range
Accuracy
Normal working condition
Storage & transportation
condition
Measurement perimeter
of the arm
Weight
External dimensions
Attachment
Mode of operation
Continuous operation
Degree of protection
Type BF applied part
Protection against
ingress of water
Software version
Device classification
Battery Powered Mode:
Internally Powered ME Equipment
AC Adaptor charged Mode: Class II ME Equipment
WARNING: No modification of this equipment is allowed.
Pressure:
5°C-40°C within ±3 mmHg (0.4 kPa)
Pulse value: ±5%
Rated cuff pressure:
0 mmHg ~ 299 mmHg (0 kPa ~ 39.9 kPa)
Measurement pressure:
SYS: 60 mmHg ~ 230 mmHg (8.0 kPa~ 30.7 kPa)
DIA: 40 mmHg ~ 130 mmHg (5.3 kPa ~ 17.3 kPa)
Pulse value: (40-199)beat/minute
About 22 cm ~ 32 cm or 22 cm ~ 42 cm
A temperature range of : +5°C to +40°C
A relative humidity range of 15% to 90%,
non-condensing, but not requiring a water
vapour partial pressure greater than 50 hPa
An atmospheric pressure range of :
700 hPa to 1060 hPa
Temperature: -20°C to +60°C
A relative humidity range of ≤ 93%, non-condensing,
at a water vapour pressure up to 50 hPa
If any abnormality arises during use, please check the following points:
NOTE: If the product still does not work, contact moon Customer Service. Under no
circumstance should you disassemble or attempt to repair the unit by yourself.
TROUBLESHOOTING
SPECIFICATIONS
Relax for a moment
and then measure again.
The monitor detected
motion,talking or the
pulse is too poor while
measuring.
Out of measurement
range
Relax for a moment.
Refasten the cuff and then
measure again. If the
problem persists, contact
your physician.
PROBLEM SYMPTOM
CHECK THIS
REMEDY
No power
Error
message
Display will not
light up.
Power is exhausted.
Charge the power
Power is low.
Warning
message
Click the ON/OFF button,
please wait for 4 seconds
to turn on the device
Don't press repeatedly
Unable to
charge
Poor contact of charging
port or the adapter is
not right
Please plug in the adapter
again and use 5V to
output the power adapter.
"Low power,
please charge
in time" is
displayed.
Low power
Charge the power
"Fault Code E01"
is displayed .
The pressure of the
cuff is excess or
air connector plus gets
loose. .
Check theair connector
plus ,
refasten the cuff .
"Fault Code E02"
is displayed .
"Fault Code E03"
is displayed .
The measurement
process does not detect
the pulse signal.
Loosen the clothing on
the arm and then
measure again.
"Fault Code E04"
is displayed .
A calibration error
occurred or out of
measurement range or
parameter error and so
on.
Retake the measurement.
If theproblem persists,
contact the retailer or our
customer service for
further assistance.
Refer to the warranty for
contact information and
return instructions.
"Fault Code OUT"
is displayed .
User manual, USB Cable,AC Adaptor
IP21 It means the device could protected against
solid foreign objects of 12.5mm and greater, and
protect against vertically falling water drops.