MONTREL TECNOLOGIA ADR Multi 4000 Скачать руководство пользователя страница 15

15

A - Single-phase connection – 1 Phase + N 1 CT.

B - Two-phase connection – 2  N 2 CTs or    
1 Phase 3 Wires. 

Figure 9-1 Single-phase connection system.

Figure 9-2 Two-phase connection.

Содержание ADR Multi 4000

Страница 1: ...OPERATION MANUAL ANDROID VERSION ADR Multi 4000...

Страница 2: ...n manual and safety standards will result in malfunctioning and or risk to the operator Symbols Utilized See explanation in the manual Alternate current Pulses Equipment protected by double insulation...

Страница 3: ...8 8 1 Operation Options 12 9 Connection System 14 10 Features 19 10 1 Start 20 10 2 ADR 21 10 3 Relat rios 27 10 4 Reports 28 10 5 Help 29 10 6 Settings 29 10 6 1 General 29 10 6 2 Parameters 30 10 6...

Страница 4: ...ng to the type of energy meter found and click on next details on page 21 9 Browse through the screens until reaching the phasor diagram and check if the observed phase shifts are correct details on p...

Страница 5: ...tion in electric power meters Analysis of energy deviation in the client s input extensions using the comparison Method input extension connection Electrical quantities metering portable multi meter G...

Страница 6: ...u have any questions please call our technical support at 55 19 3861 3070 or send an email to montrel montrel com br Preventive maintenance of the devices is unnecessary Corrective maintenance if nece...

Страница 7: ...ions or situations not provided for in this manual If you have any questions please call our Techni cal Support at 55 19 3861 3070 or send an email to montrel montrel com br 6 STORAGE AND OPERATION Tr...

Страница 8: ...operation indication in a Smart phone though without losses or changes for operation by other devices Figure 8 1 shows the application screen in a tablet and a smartphone Figure 8 1 Operation screen...

Страница 9: ...a gauging box if the strap is facing backwards More details in figure 8 3 where shows the assembly for hanging it on the input extension and shows the assembly for fixing it to a panel door or the edg...

Страница 10: ...set of 200A clamps and a rate of 7 200 000 impulses kWh or kvarh to the set of 10A clamps configurable to facilitate ADR Multi 4000 calibration Table 8 2 shows the electric functions of the output co...

Страница 11: ...play of the pins and the characteristics of the external trigger and pulse input connector pino do Item shows the input of signals from the current clamps It is possible to connect up to four clamps t...

Страница 12: ...r Other options may be provided according to the need and upon business proposal Voltage signal connection uses the crocodile clips seen in figure 8 10 Figure 8 10 Voltage Connections Figure 8 11 Arti...

Страница 13: ...s in which the sensor is fixed to the front part of the meter through suction cups and can detect the disk mark or the LED pulse this sensor is shown in figure 8 13 in the second option should it be n...

Страница 14: ...of the related values such as the ener gy consumed and the power among others that are indispensable to inspection quality control and identification of failures and issues in electric power distribut...

Страница 15: ...15 A Single phase connection 1 Phase Neutral 1 CT B Two phase connection 2 Phases Neutral 2 CTs or 1 Phase 3 Wires Figure 9 1 Single phase connection system Figure 9 2 Two phase connection...

Страница 16: ...C Balanced three phase connection 1 Phase Neutral 1 CT D Three phase connection without neutral 3 Phases 2 CTs Figure 9 3 Balanced three phase connection Figure 9 4 Three phase connection without neu...

Страница 17: ...7 E Three phase connection without neutral 3 Phases 3 CTs F Three phase connection 3 Phases Neutral 3 CTs Figure 9 5 Three phase connection without neutral with 3 CTs Figure 9 6 Three phase connection...

Страница 18: ...LTI4000 G Indirect three phase connection 3 Phases Neutral 3 CTs H Indirect three phase connection 3 Phases 2CTs Figure 9 7 Indirect three phase connection Figure 9 8 Indirect three phase connection 2...

Страница 19: ...optional serving as an indication of the actual value of the current in the neutral conductor Therefore the energy parameters as well as powers do not take this parameter into account 10 FEATURES Thi...

Страница 20: ...conducted in the power meters has access to the Reports saved in the device and to the Help module which enables the user to refer to the operation manual of the device After accessing the applicatio...

Страница 21: ...actors through data organized in tables charts and phasor dia grams which enable the user to conduct a quick and clear analysis of the behavior of the network connected to the meter being inspected ea...

Страница 22: ...uit Indirect ADR connected to the circuit through CT and TP A constant that makes a relation between the energy measured and the number of turns pulses of the meter Imp kWh or Wh Imp No of turns pulse...

Страница 23: ...ded in a table Instant values by phase recorded in a linear chart Instant values by phase re corded in a phasor diagram and tables Table 10 2 Metering organization After properly configuring the param...

Страница 24: ...ustrates the tab indicating the averages and totals measured quanti ties Figure 10 5 Tab screen of Averages and Totals Figure 10 6 Tab screen of instant measures by phase Figure 10 6 shows the tab scr...

Страница 25: ...Figure 10 7 Linear chart screen Figure 10 8 Phasor diagram screen Figure 10 7 shows the instant visualization screen in the form of linear charts Finally figure 10 8 illustrates the phasor diagram ta...

Страница 26: ...to the metering con stant which makes a relation between the measured energy and the number of turns pulses of the meter In Pulses Turns counting a pulse is given in the application every time the me...

Страница 27: ...to view the completed tests saved in the ap plication It is also possible to print reports in the field save in pdf or export the test data in csv or html files The application does not limit the numb...

Страница 28: ...user profile WARNING This procedure changes the device s internal parameters and should only be performed by a duly qualified person To perform calibration select the ADR to be calibrated through the...

Страница 29: ...mpatibility of the Android versions For more information please contact the Montrel team 10 6 SETTINGS In the configuration screen the user can adjust the application functions as needed The configura...

Страница 30: ...nclude the logo in the report click once on the Report logo item and then upload the image file containing the company logo The image file extension must be jpg image resolution must be 300 dpi and it...

Страница 31: ...lt screen In the event that a number of 0 cycles is selected the device performs the testing cycles on a continuous basis until the user ends the test and subsequently displays the screen with the res...

Страница 32: ...mation on the meter inspection must be displayed in the test configuration in the result of the tests and in the inspec tion report Figures 10 16 10 17 and 10 18 show an example of the three configu r...

Страница 33: ...33 Figure 10 18 Test report module Table 10 3 provides the description of each parameter subject to configuration by the user...

Страница 34: ...S U U L L T T I I N N F F O O I I n n s s p p e e c c t t i i o o n n I I D D Report No Enables display of the inspection report No ID on the test result page Date and Time Enables display of the ins...

Страница 35: ...irect TP Ratio Enables display of the identification of the voltage transformation ratio considered on the test result page in case of indirect metering CT Ratio Enables display of the identification...

Страница 36: ...testing on the test result page Approval Enables display of the approval of the inspected meter on the test result page Approved Rejected Estimated Constant Enables display of the estimated constant f...

Страница 37: ...ody of the report Service Order Enables display of the inspection service order ID on the body of the report Client Code Enables display of the client code ID on the body of the report Consuming Unit...

Страница 38: ...nsidered meter class on the body of the report Class A 2 Class B 1 Class C 0 5 Class D 0 2 Meter Constant Enables display of the considered metering constant on the body of the report No of turns or p...

Страница 39: ...er on the body of the report Total apparent power kVA Enables display of the total measured apparent power on the body of the report Total power factor Enables display of the total power factor calcul...

Страница 40: ...R Multi 4000 serial No on the body of the report Calibration report No Enables display of the ADR Multi 4000 calibration report No on the body of the report Date of Calibration Enables display of the...

Страница 41: ...perform tests check reports and support documentation as well as to prevent access by the user to the use settings To cre ate a new profile click on the New icon located on the lower right corner of t...

Страница 42: ...hers information on each in dividual or group of individuals that will operate ADR Multi 4000 The user is identi fied with a registration name and number This is an optional number to identify the ope...

Страница 43: ...en be opened and the name of the new user the profile to which the user will be linked the user registration number and the password to access the application must be provided in it To complete the cr...

Страница 44: ...e and yellow 2 0m CAT IV 300V 04 Testing leads 04 Alligator clips black red blue and yellow 04 Power clips Transportation and storage case Operation manual in English Safety guidelines Optional parts...

Страница 45: ...her warranty is implicit or explicit and therefore the manufacturer is not liable for any loss or damage directly or indirectly caused by misuse of the device for the intended purpose This warranty wi...

Страница 46: ...46 ADR MULTI4000 NOTES...

Страница 47: ...47 NOTES...

Страница 48: ...Innovation that is our differential 4 4 45 Maria Caporali St Mogi Guacu SP Brazil Zip Code 13845 021 55 19 3861 3070 montrel montrel com br www montrel com br...