6
SERVICING
An annual service is recommended to maintain optimum operating conditions and will
include the following points:
•
Check / replace pre-filter
•
DOP test the main HEPA filter if required
•
Particle count check if required
•
Check airflow monitor and re-calibrate if necessary
•
Check and record air velocity readings
•
Check general condition of system.
•
Issue test report and airflow certificate.