background image

 

4

INSTALLATION AND SETUP 

 
The module is freestanding and therefore requires no special mounting for it’s 
operation. It should be positioned to allow unobstructed airflow into the prefilter. 
 

SET UP FAN SPEED

 

This requires the use of a calibrated Ø100mm rotating vane anemometer with 
readings taken from 150mm from filter face. 
 
Turn the system on and set the fan speed using the speed control knob. Set to the 
required level using the anemometer to measure the air velocity from the HEPA filter. 
 
The air volume required to achieve N air changes per hour can be calculated by the 
following formula: 

Air volume required (m³/hour) = Room Volume (m³) X N 

 
The air volume through the module can be calculated by the following formula: 
 
 
 

CAM1500F

 : Air Volume (m³/hour) = 0.595 (filter area in m²) X measured HEPA 

velocity (m/sec.) x 3600 (secs/hr) 
 
Based on the filter size used in this module (850mm x 700mm), the following table 
gives an indication of air volume (m³/hour) against HEPA velocity: 
 
HEPA Velocity (m/sec.) 

Air Volume (m³/hour) 

0.3 

643 

0.35 

750 

0.4 

857 

0.45 

964 

0.5 

1071 

0.55 

1178 

0.6 

1285 

0.65 

1392 

0.7 

1500 

0.75 

1606 

 
 

 

Please note: 
This unit is capable of processing air volume is excess of 1500m³/hr but it 
should be noted that the filter will be potentially stressed beyond it’s design 
limits, filter life will be greatly reduced and noise levels will increase 
substantially. 
It is therefore advised that the unit is not operated beyond 0.7m/sec 
(1500m³/hr). 
 
 
 

Содержание CAM-F 1500

Страница 1: ...Operating Maintenance Manual www monmouthscientific co uk Clean Air Module CAM F 1500 ...

Страница 2: ... in any form or by any means redistributed or reproduced in entirety or in part in any form for individual or third party use photocopied or stored on any form of electronic retrieval system or linked to any other website without the prior express written permission of Monmouth Scientific Limited For permission requests please write to Monmouth Scientific Limited Units 5 5 Kilnside East Quay Bridg...

Страница 3: ... self contained filtration unit designed to fit into a wall to provide HEPA filtered clean air The module requires a 230vac 3A electrical supply supplied via a domestic 13A socket outlet Speed control knob Power inlet socket c w fuse holder 3A fuse Power switch ...

Страница 4: ...red m hour Room Volume m X N The air volume through the module can be calculated by the following formula CAM1500F Air Volume m hour 0 595 filter area in m X measured HEPA velocity m sec x 3600 secs hr Based on the filter size used in this module 850mm x 700mm the following table gives an indication of air volume m hour against HEPA velocity HEPA Velocity m sec Air Volume m hour 0 3 643 0 35 750 0...

Страница 5: ...securing the pre filter retaining frame 6 Reconnect the mains supply and switch on 7 Check the airflow and adjust if necessary as detailed in the setup section CHANGING HEPA FILTER 1 Turn the system off and disconnect from mains 2 Remove the screws securing the filter retaining frame 3 Remove the filter frame and HEPA filter 4 Replace HEPA filter with the correct type see page 5 5 Replace cover fi...

Страница 6: ...k replace pre filter DOP test the main HEPA filter if required Particle count check if required Check airflow monitor and re calibrate if necessary Check and record air velocity readings Check general condition of system Issue test report and airflow certificate For parts or service information please contact Monmouth Scientific ...

Страница 7: ...7 ...

Страница 8: ...t i s h D e s i g n Manufacture Monmouth Scientific Limited UK Headquarters Units 5 6 Kilnside East Quay Bridgwater Somerset TA6 4DB 44 0 1278 458090 www monmouthscientific co uk info monmouthscientific co uk ...

Отзывы: