ELECTRICAL INSTALL ATION

SETUP

      

Hazardous Location Precautions:

Observe the regulations listed in the NEC (USA) or CEC

(Canada)  regarding  equipment  in  hazardous  locations

when  installing  the  hazardous  location  models  of  the

Vibra

Rod

™  level  sensor  in  hazardous  classified  loca-

tions. Wiring must be performed according to the authority

having site jurisdiction and is the sole responsibility of the

installer to ensure such.

It is important to note that the hazardous location mod-

els are certified with two protection concepts: The probes

and wires from them to the enclosure is an intrinsically-

safe circuit but is still Haz-Live. The enclosure and elec-

tronics within are not intrinsically-safe and protection is by

dust-ignition proof enclosure.

Factory Wiring:

The probe wires internal to the housing are connected to the

frontside of the PCB.  DO NOT alter this connection.  Doing so

will likely cause improper operation of the sensor and likely per-

manent damage. (See Figure 7)

      

Permanently Connected Equipment:

Disconnecting  devices  shall  be  included  in  the  system

installation.    In  installations  where  multiple  circuits  are

used (i.e. independent circuits for power input and relay

output), individual disconnects are required.  The discon-

nects  shall  be  within  close  proximity  of  the  equipment,

accessible to operators, and marked appropriately as the

disconnect for the associated circuit.  Assure the discon-

nect ratings are appropriately sized to the circuit protected

(See Specifications). The MAINS power that connects to

the  sensor  MUST  provide  circuit  protection  and    NOT

exceed 30A total.

Circuit Separation 

(Standard version only)

:

Two cable entry locations are provided to aid in maintain-

ing  separation  of  “hazardous  live”  (typically  mains  volt-

ages such as 115VAC and 230VAC) and limited circuits

(typically control voltages less than 30Vrms or 42.5VDC).

However, since the 

Vibra

Rod

sensor’s single wiring com-

partment can not absolutely protect against physical con-

tact between multiple circuits, it is required that all wiring

used  must  have  an  insulation  rating  of  300V  minimum,

and a temperature rating of 176˚ F (80˚ C) minimum.

Protective Earthing:

Each 

Vibra

Rod

level sensor is provided with a “protective

conductor terminal” which shall be terminated to the local

earth  ground  potential  of  the  facility  to  eliminate  shock

hazard  in  the  unlikely  event  of  internal  insulation  break-

down.  Select wire size that can carry in excess of the sum

of all circuits maximum amperage.

1)

 Remove the housing cover to access the terminal 

    blocks and operating mode switches. All wires must be

    routed through the conduit openings.

2) Power Input:

(See Figure 7) The 

Vibra

Rod

sensor is 

    provided with a HI / LOW voltage, AC/DC universal 

    power supply.  If using a DC power supply, polarity is 

    irrelevant.  

3)

 Do not remove the green wire from J1 MAINS terminal

    because it is an internal connection.  For grounding the

    unit, either use the grounding screw terminal on the 

    outside of the housing or use the internal grounding 

    screw terminal on the circuit board (J1 MAINS). The 

    earth ground screw should be connected to a quality 

    "ground" to eliminate shock hazard.

4) Output Contact Connections:

The SPDT relay con-

    tacts can be used to indicate whether or not material is

    being detected by the sensor within the vessel.  These

    outputs are also influenced by the setting of the Fail-

    Safe switch (See Setup).  The output contact(s) can 

    be used to switch high voltage/high current loads as 

    listed in the Specifications.  Whenever possible, use an

    independent voltage source to operate the loads.  

    Select wire size that can deliver suitable voltage and 

    current for the application.  Follow all electrical codes 

    and use proper wire gauge size.  Tighten cable glands

    as well as housing cover after installation to ensure 

    proper sealing for “Type 4X” environmental protection.
    

Conduit Entries

: Conduit entry threads should be 

    lubricated with a material such as lithium grease to 

    ensure water ingress protection required to maintain 

    the Enclosure “Type 4X” rating. 

-

/2:

)

6

+,

& % $

6(

1

6

-

)

-

-

-

/1(8

1&&2012

67$786/('V

)$,/6$)(

6(/(&725

6(16,7,9,7<

6(/(&725

)$725<352%(

&211(&7,21

'2127$/7(5

)$&725<&211(&7,21

'2127$/7(5

Figure 7

Complete set-up of the 

Vibra

Rod

™ is accomplished sim-

ply by adjusting the sensitivity and fail-safe settings.  Each

application  can  be  adjusted  independently  for  optimum

operation.

Material Density:

The SENSITIVITY selector should be set in accordance

with the density of the target material:
Position A, high sensitivity. For materials with density over

10.0~12.5Lbs/Ft

3

(160~200kg/m

3

).  Recommended  for

loose and light materials, greatest sensitivity of detection.
Position B, medium sensitivity position: For materials with

density 12.5~15.6Lbs/ft

3

(200~250kg/m

3

). Recommended

for modest density materials.
Position C, low sensitivity position: For materials with den-

sity 18.7~21.8Lbs/ft

3

(300~350kg/m

3

). Recommended for

thick and heavy materials, least sensitivity of detection.
If  the  material  to  be  sensed  can  potentially  form  heavy

deposits  on  the  probe,  selection  "LOW  sensitivity"

(Position “C”) should be chosen.  This will provide addi-

tional immunity to product buildup while still enabling the

circuit to sense the presence of material.

AC

DC   +   -

FACTORY PROBE

CONNECTION

(DO NOT ALTER)

Содержание DuraVibe VibraRod

Страница 1: ...ertaining to SAFETY Doing so will ensure proper operation of the equipment and the safety of all personnel Before discarding shipping container please inspect it thoroughly and verify that all parts o...

Страница 2: ...electing a location for the VibraRod level sensor choose a point in the vessel where the probe will be out of the direct flow of incom ing and outgoing material to prevent any mechanical damage that m...

Страница 3: ...ING INSIDE EXTENSION PIPE ELECTRI DO NOT DRILL OR WELD THE DEVICE FOR SECURING INSIDE EXTENSION PIPE ELE DO NOT DRILL OR WELD THE DEVICE FOR SECURING INSIDE EXTENSION PIPE ELEC DO NOT DRILL OR WELD TH...

Страница 4: ...cover to access the terminal blocks and operating mode switches All wires must be routed through the conduit openings 2 Power Input See Figure 7 The VibraRod sensor is provided with a HI LOW voltage...

Страница 5: ...observing LED s GREEN not sensed RED sense no light no power 3 Verify sensitivity setting Position sensitivity selection in position Highest Sensitivity A thereby making the probe more sensitive to di...

Страница 6: ...DV L00 STANDARD PROBE 07 19 18 B ADV L006 STANDARD PROBE 07 19 18 B ADV L0064 3 STANDARD PROBE 07 19 18 B ADV L0064 3 STANDARD PROBE 07 19 18 B ADV L0064 3 STANDARD PROBE 07 19 18 B ADV L0064 3 5 STAN...

Страница 7: ...r of any defects within the warranty period return the product intact and prepay transportation charges The obligation of Monitor Technologies LLC under this warranty is limited to repair or replaceme...

Страница 8: ...Monitor Technologies LLC 44W320 Keslinger Rd t Elburn IL 60119 t 630 365 9403 t 800 766 6486 t monitor monitortech com t www monitortech com B U L L E T I N 554A JSC JSC JUL 20 2018 DOC 001 554A 2...

Отзывы: