14 

The Ethernet Connection is an option and will be installed only if specifically ordered. It is 
a 10MHz TCP/IP port that allows the user to program the recorder remotely, start or stop 
recording and download real time data. 
 
Refer to the separate manual that is supplied with the ethernet interface option. 

 

2.3.6 USB Port (Option) 

 

The USB Port is an option and will be installed only if specifically ordered. It is a serial 
connection to a PC that allows the user to program the recorder remotely, start or stop 
recording and download real time data. 
 
Refer to the separate manual that is supplied with the USB interface option. 

 

2.3.7 Cleaning 

 

The unit may be cleaned by wiping with a soft cloth. The front panel and display / keypad 
may be wiped with a slightly damp soft cloth containing soapy solution or a mild 
detergent. 

Do not use any lemon based (citric acid) product to clean the display / 

keypad. NOTE: THE UNIT IS NOT WATERTIGHT.

 

 

2.4 INPUT 

SCALING 

 
The input to the recorder is often a standard process variable (e.g. voltage) rather than the 
actual units being measured (e.g. pressure).  
 
There are a number of custom conversions available in the unit to take care of special inputs 
such as thermocouples and RTDs, which need special scaling and linearization. These are 
preset in the unit and require no modification. There is also the capability to customize any 
input variable that has a linear relationship to the measured variable. The relationship 
between the required variable and the input variable must satisfy the constraint 
 
 

 

y = mx + c

 

 
where 

y

 is the value to be displayed and recorded, 

x

 is the input variable (typically volts or 

milliamps DC), 

m

 is the scale factor which defines the relationship between the input and the 

displayed variable and is a constant (linear relationship), and 

c

 is a constant offset variable 

that may be positive or negative. 
 
This facility to scale the input is found in the ADVanced SETUP menu and is available to 
each channel independently in dual channel units. Here the user can easily scale the input 
and set the displayed units using three alphanumeric characters. 
 
The scaling is achieved in the unit in the CHANnel X SETUP option (X = A or B) in the 
ADVanced SETUP menu. First set the displayed LO SCALE, which may be positive, negative 
or zero. Then set the displayed HI SCALE for the full-scale input. The unit will then 

automatically compute the scale factor and offset.   
 
Setting the engineering units is done in the same menu 
location using the CHAN UNITS option to set the three 
alphanumeric characters.  
 
Let us assume we have an input from a pressure 
transducer of 1 to 5 volts DC, corresponding to real 
world units of zero to 250 PSI (Pounds per Square Inch 
of pressure). Assume we choose the 0 to 5 volt input. 
The actual output span of the transducer is 4 volts, with 
1 volt = 0 PSI and 5 volts = 250 PSI. This is equivalent 
to 62.5 PSI per volt. The LO SCALE which is the "0" volt 

5

4

3

2

1

0

0

125

250

-62.5

(OFFSET)

INPUT

VOLTS

PSI

Содержание Data Chart 1250

Страница 1: ......

Страница 2: ...igital Input 2 3 5 Ethernet Connection Option 2 3 6 USB Port Option 2 3 7 Cleaning 2 4 INPUT SCALING 2 5 ALARMS 2 5 1 Alarm Types 2 5 2 Setting Alarms 2 5 3 Alarm Indication 2 6 MEMORY CARDS 2 6 1 Car...

Страница 3: ...DS AND PRECAUTIONS Read and follow all instructions in this manual carefully and retain this manual for future reference Do not use this instrument in any manner inconsistent with these operating inst...

Страница 4: ...eciclar contacta el distribuidor para detalles Afin de respecter la directive europ enne 2002 96 EC sur l Equipement Electronique et les D chets Electriques Ce produit pourrait contenir des mati res q...

Страница 5: ...ich the samples appear to move across the Graphics Display but unlike paper recorders the display is made up of discrete data points each of which is a distinct sample of the data being measured The r...

Страница 6: ...the average or detect and store either the maximum peak value or minimum valley value depending on the mode programmed It will then store this value when the minute is up Unless required otherwise it...

Страница 7: ...tion Can also be used as a means of updating the firmware in the unit Cards up to 2 Gigabytes in size can be used B Busy LED Red on when the unit is recording and blinks whenever the unit is accessing...

Страница 8: ...ight arrow buttons The values in the digital window J are the values at the point the cursor intersects the trend line L Date Time Axis Delimiter dashed vertical lines placed at equal increments along...

Страница 9: ...l mode as shown below enables both channels to be shown as Digital values similar to a digital panel meter The user has the option of showing the displays in normal or reverse video shown modes and th...

Страница 10: ...he factory for repair contact the sales organization from which you purchased the instrument A Return Merchandise Authorization RMA number must be obtained from the factory before returning an instrum...

Страница 11: ...Cut a panel opening 96 mm x 96 mm 3 62 x 3 62 in 2 2 1 2 Remove any packaging material from the Recorder Always handle the unit carefully to avoid damaging the LCD display or scratching the display s...

Страница 12: ...Before connecting any power or inputs to the unit ensure that all signal wires and power cables are free of potential The unit itself is powered from 9Vdc 0 5V 4 5VA as standard although other option...

Страница 13: ...ollow the same procedure carefully identifying the positive supply line The unit is protected against reverse polarity but not against substantial over voltage 2 3 2 Input Signal Connections The maxim...

Страница 14: ...are used the shield should be grounded at the sensor end only 2 3 3 Relay Output Connections To prevent the possibility of electrical shock use extreme caution when wiring contact output connections H...

Страница 15: ...has a linear relationship to the measured variable The relationship between the required variable and the input variable must satisfy the constraint y mx c where y is the value to be displayed and re...

Страница 16: ...to 5 and the alarm is a high type and non latching then the alarm will activate when the input exceeds 100 and will remain active until the input drops below 95 100 minus 5 Alternatively the user can...

Страница 17: ...ed in alarm condition the status indicator will blink and the internal beeper will sound The beeper can be permanently disabled from the Advanced Setup Menu BEEPER ALARMS OFF The relays close on an al...

Страница 18: ...recording The files on the card are in DOS standard format and can be managed on any Windows based machine with a CompactFlash reader internal or external The content of the files is in proprietary b...

Страница 19: ...memory card or from the internal buffer and is available as far back as the start of recording When fast forwarding data is available up to real time The time and date stamp on the screen indicates th...

Страница 20: ...ny part of the graph within the zoom window Once in a zoom level all functions are operational on the zoomed display When in a zoom mode the zoom level is indicated by a icon on the status line No ind...

Страница 21: ...RD MODE menu option may be password protected Before selecting ON or TRIGGERED to begin a recording ensure that a Memory Card is in place and that there is sufficient space available to contain your n...

Страница 22: ...on the prerecorded file Use the Left or Right buttons to move through the file To return to the current real time trend press the MENU button 3 5 6 CARD STATUS To display the status of the memory card...

Страница 23: ...nally which will enable it to automatically change sampling speed or begin or end a recording session The trigger can be any alarm or the external input The user can thus set the unit to record at a s...

Страница 24: ...press the MENU button This symbol indicates the display is in the COMPRESSED mode The amount of compression can be determined from the time date stamps on the display To exit from the Compressed Mode...

Страница 25: ...ist directory Press MENU to save and exit Unit will restart on successful load CURSOR ID Puts chart in HOLD mode and places cursor on screen Use keys to move cursor and read trend values at top Press...

Страница 26: ...channel for stability of readings or to catch peak values if the display rate is slower than the sample rate The unit always samples at 100 samples per second per channel For increased accuracy each r...

Страница 27: ...arm till relays pull in RESET DELAY Use keys to enter time in seconds from RESET till relays drop out CHAN TAG Enter a Channel Tag Identifier up to 5 Characters Use keys to clear or escape Use MENU ke...

Страница 28: ...lsafe BOTH Select both relays as fail safe 4 4 DISPLAY This option is used to set up the display The Window Configuration allows the user to choose how to display data either as a trend or as digital...

Страница 29: ...AN B TREND line dashed instead of solid 4 5 DATA CARD This option contains duplicate menu items to those found in the RECORD MODE menu In addition it enables the user to test the card and format it fo...

Страница 30: ...is MM DD YY SET TIME Set Time Adjusts hours adjusts minutes MENU accepts SET DATE Set Date Adjust month year adjusts days MENU accepts 4 7 BEEPER This option allows the user to have various events so...

Страница 31: ...TRIG and press the MENU button Select either OFF ALARM 1 through 4 or EXTERNAL INPUT as the record trigger Note that if an alarm reset delay is set see Section 4 2 the unit will continue to record fo...

Страница 32: ...a span correction value above or below actual 1 also in small amounts NOMINAL depending on the input range selected either resets the correction values to 0 and 1 or copies those values from the range...

Страница 33: ...bed above Then switch to Thermocouple range Select NOMINAL to copy the volts calibration to thermocouple range The recorder is now calibrated for ideal sensors Refine the calibration by either shortin...

Страница 34: ...l and then copying the calibration values using NOMINAL should be enough Remaining sources of error are the ambient sensor probably small for the thermocouple and the current source small for the RTD...

Страница 35: ...to 1000 C 2 C 148 to 1830 F 3 F T 100 to 400 C 2 C 148 to 750 F 3 F E 80 to 400 C 2 C 112 to 750 F 3 F RTD 2 or 3 wire Resolution 0 1 C Accuracy 0 3 Internal current source 1mA 100 ohm Pt385 100 to 75...

Страница 36: ...1010 1 Altitude up to 2 000 m Temperature 10 C to 50 C operating per IEC61010 1 Humidity Maximum relative humidity 80 for temperatures up to 31 C decreasing linearly to 50 relative humidity at 40 C Ce...

Страница 37: ...ho has the right password BAD FILENAME The filename entered does not meet the DOS requirement Use A to Z and 0 to 9 without spaces 8 charactres max Re enter the name CARD ABSENT Failed attempt to read...

Страница 38: ...wn simultaneously The DOWN and LEFT arrows while powering up the unit Wait two to three seconds and then release APPENDIX D RE PROGRAMMING FIRMWARE The operating firmware in the unit is programmed int...

Отзывы: